Beckman Coulter Inc.: AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.

Recall number
Z-3135-2017
Recalling firm
Beckman Coulter Inc.
Product
AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.
Status
Terminated
Initiated
2017-08-21
Posted
not on file
Terminated
2023-09-27
Reason
Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which manifests in incorrect values being reported out of the laboratory for some patient samples.
Root cause
Under Investigation by firm
Action
The firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letter dated August 16, 2017 to the affected customers via email, mail, and phone during the week of August 21, 2017. The letter described the product, problem and actions to be taken. The customers are instructed to follow instructions provided in the letter and complete and return the Customer Response Form within 10 days via Fax to: (786)-639-7500 /7501 /7502 /7504; mail to: Beckman Coulter, Inc.,- Attn: Regulatory Affairs Mail Code 31-B06; 11800 S.W. 147th Avenue; P O BOX 169015, Miami, FL 33116-9015 or email (if received electronically) at: Regulatory.Notifications@beckmancoulter.com -include in subject field: FA-31696 AQUIOS CL August 2017. If you have any questions regarding this notification, please contact: From our website: http://www.beckmancoulter.com By phone: call 800-369-0333 in the United States and Canada. By email: LScustomerLetter@Beckman.com Outside the United States and Canada, contact your local Beckman Coulter representative.
Quantity
213 units (34 units in the US)
Distribution
Worldwide Distribution: United States (nationwide) and countries of: Angola, Australia, Botswana, Canada, Chile, Columbia, Ghana, Israel, Malaysia, Namibia, Pakistan, Panama, South Africa, Uganda, France, Switzerland, Poland, Portugal, Germany, Belgium, Finland, Czech Republic, Denmark, Netherland, Turkey, Spain, Great Britain, Romania, Brazil, Italy, Sweden, Ghana, Colombia, Lesotho and Zambia
Lot, serial or model codes
All software versions.; Serial numbers - BA30048, AY20032, BA01007, AZ19046, AZ12030, BA25046, BA05011, AY51072, BA01002, AZ15044, BA30049, BA30050, BA05015, AZ19047, BA18029, AY47071, AZ15038, BA23036, BA23035, AZ23058, AZ20057, AY47063, BA15024, BA15023, AZ45080, BA01010, AZ25070, AZ32071, AZ12033, AZ25065, AZ25068, AY51075, AZ19052, BA15025
510(k) numbers
["K141932"]
PMA numbers
not on file
Product code
OYE
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register