Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: FFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLab POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable

Recall number
Z-3132-2017
Recalling firm
Boston Scientific Corporation
Product
FFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLab POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable
Status
Terminated
Initiated
2017-05-15
Posted
not on file
Terminated
2021-07-06
Reason
The device history record (DHR) was missing its test documentation for final HIPOT (high potential) electrical testing.
Root cause
Unknown/Undetermined by firm
Action
Boston Scientific sent an Important Medical Device Information letter dated May 30, 2017, to their affected consignee. On May 15, 2017, Boston Scientific visited the consignee and exchanged the affected device for a device that met specifications. The letter requested a response form be completed and returned. For questions customers should call 763-494-1133.
Quantity
3
Distribution
US Distribution to one customer in Missouri.
Lot, serial or model codes
Lot/Batch No. (Exp Date): SPM01975 (09/05/2044), SPM00890 (09/27/2043), SPM01616 (06/28/2044)
510(k) numbers
["K170204"]
PMA numbers
not on file
Product code
DRG
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.