Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Inhibin A ELISA, Part Number: DSL-10-28100-4 Product Usage: The Inhibin A ELISA kit is intended for the quantitative measurement of dimeric Inhibin A in human serum or plasma. It is intended strictly for in vitro use as an aid in the diagnosis and monitoring of various hormonal reproductive disorders.

Recall number
Z-3128-2011
Recalling firm
Beckman Coulter Inc.
Product
Inhibin A ELISA, Part Number: DSL-10-28100-4 Product Usage: The Inhibin A ELISA kit is intended for the quantitative measurement of dimeric Inhibin A in human serum or plasma. It is intended strictly for in vitro use as an aid in the diagnosis and monitoring of various hormonal reproductive disorders.
Status
Terminated
Initiated
2010-11-10
Posted
2011-09-02
Terminated
2012-07-11
Reason
The recall was initiated because Beckman Coulter has confirmed the microbial contamination in the conjugate diluent bottles included in the Manual Inhibin A Elisa kit (Part Number DSL-10-28100-4), kit lot number 110761.
Root cause
Other
Action
Beckman Coulter initiated an "URGENT PRODUCT CORRECTION" on June 15, 2011, via a telephone script to the only one customer impacted by the recalled product. The telephone script provided an explanation of the problem identified and the actions to be taken. Beckman Coulter recommended that the customer discontinue use of the recalled product as a preventative measure. Beckman will send a representative to the customer's facility to replace the recalled product in all remaining kits of the affected lot. The customer was instructed to share this information with laboratory staff. For any questions concerning this information, contact Beckman Coulter at 952-368-7874.
Quantity
342 units
Distribution
(USA) Nationwide Distribution in the state of CA
Lot, serial or model codes
Lot # 110761
510(k) numbers
not on file
PMA numbers
not on file
Product code
NDR
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register