Skip to the content

FDA Medical Device Recalls (openFDA)

Hitachi America, Ltd., Particle Therapy Division: Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. Component: N/A

Recall number
Z-3125-2024
Recalling firm
Hitachi America, Ltd., Particle Therapy Division
Product
Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. Component: N/A
Status
Open, Classified
Initiated
2024-05-27
Posted
2024-09-12
Terminated
not on file
Reason
During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.
Root cause
Software design
Action
Between 08/05-09/2024, the firm communicated/hand-delivered an "URGENT: MEDICAL DEVICE RECALL" letter to customers informing them that during positioning in the Positioning Image Analysis Systems (PIAS) 3D3D matching mode, while transferring Cone Beam CT systems (CBCT) images from CT Image Reconstruction Software (CIRS) to PIAS, when the Float Image Selection button on the PIAS screen is pressed, it may display CBCT images that are insufficient in number. This may reveal a clear displacement in the Y-direction (body axis direction) compared to the planning CT. There may potentially be a possibility of completing 3D3D patient positioning approximately 5mm off in the body axis direction (with 5 images missing) and proceeding with treatment. Customers are instructed, until the software has been replaced with the corrected version, all users of the device are directed to follow the interim countermeasure. When users operate as described in the operation manual, no such event should occur. Countermeasure references: Engineering Document: PBT-40E-1042 Instruction for patient positioning assistance using CBCT [PROBEAT-FR] Operation Manual TS.16F0147-1 Page.1-105 / 1-108 [PROBEAT-CR] Operation Manual TS.16F0147-1 Page. 1-4-20 / 1-4-23 [PROBEAT-V] Operation Manual TS.16A1850-1E Page.1-126 / 1-131 For questions/assistance -
Quantity
1
Distribution
U.S. Nationwide distribution in the states of TN, TX, and WA.
Lot, serial or model codes
Lot Code: UDI:(01)04560333350204(11)240112(21)PTFR-01
510(k) numbers
["K232032"]
PMA numbers
not on file
Product code
LHN
Firm FEI
3003993895
Firm city
Santa Clara
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register