Beckman Coulter Inc.: Unicel DxH 800 Coulter Cellular Analysis System Product Usage: UniCel DxH 800 Coulter Cellular Analysis System Instructions for Use PN 629743AD The UniCel DxH 800 Analyzer is a quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel DxH 800 Analyzer provides a: 1. Complete Blood Count (CBC), Leukocyte 5 Part Differential (Diff), Reticulocyte (Retic) and Nucleated Red Blood Cell (NRBC) on whole blood. 2. Total Nucleated Count (TNC) and Red Cell Count (RBC) on Body Fluids (cerebrospinal, serous and synovial) (BF).
- Recall number
- Z-3125-2011
- Recalling firm
- Beckman Coulter Inc.
- Product
- Unicel DxH 800 Coulter Cellular Analysis System Product Usage: UniCel DxH 800 Coulter Cellular Analysis System Instructions for Use PN 629743AD The UniCel DxH 800 Analyzer is a quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel DxH 800 Analyzer provides a: 1. Complete Blood Count (CBC), Leukocyte 5 Part Differential (Diff), Reticulocyte (Retic) and Nucleated Red Blood Cell (NRBC) on whole blood. 2. Total Nucleated Count (TNC) and Red Cell Count (RBC) on Body Fluids (cerebrospinal, serous and synovial) (BF).
- Status
- Terminated
- Initiated
- 2011-05-09
- Posted
- 2011-09-01
- Terminated
- 2012-06-18
- Reason
- The recall was initiated because Beckman Coulter has confirmed that the probe wash collar tubing may become stretched or trapped at the Sample Aspiration Module (SAM) during routine use of single-tube presentation station. When the tension on the tubing is severe, the tubing may not be able to return to its home state and may pull the wash collar out of alignment with the aspirate probe. The misal
- Root cause
- Pending
- Action
- Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated May 11, 2011 with an attached Product Correction Response Form to all affected customers. The letter provides the customers with an explanation of the problem identified, potential impact should the problem occur and actions to be taken. The letter states that this issue will be corrected in a future hardware release. Customers were instructed to share this information with laboratory staff and retain this notification as part of Quality System documentation. If ownership or location of the analyzer(s) has been transferred to another laboratory, customers were instructed to provide a copy of the letter to that party. Customers were also asked to complete and return the enclosed Response Form within ten (10) days, so Beckman can be assured they have received this important communication. Questions concerning this notice, were directed to the Beckman Coulter Customer Service at 800-526-7694 in the United States or Canada, or contact their local Beckman Coulter representative.
- Quantity
- 680 units
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Australia, Belgium, Canada, Colombia, Croatia, Czech Republic, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Korea, Republic of Macao, Malaysia, Myanmar, Netherlands, Norway, Philippines, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
- Lot, serial or model codes
- Part Number: 629029
- 510(k) numbers
- ["K081930"]
- PMA numbers
- not on file
- Product code
- GKZ
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28