Skip to the content

FDA Medical Device Recalls (openFDA)

C.R. Bard Inc: Products that contain the Pericare Wipes: SureStep" Foley Tray System Add-A-Foley Catheter REF: A153200 UDI code: 00801741121067 10801741121064 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A304400A UDI code: 00801741073847 10801741073844 SureStep" Foley Tray System Add-A-Foley Catheter REF: A399400 UDI code: 00801741100369 10801741100366 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A399400A UDI code: 00801741073915 10801741073912 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A904400A UDI code: 00801741074097 10801741074094 SureStep Intermittent Catheter Tray REF: INTAAI UDI code: 00801741183317 10801741183314 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.

Recall number
Z-3118-2026
Recalling firm
C.R. Bard Inc
Product
Products that contain the Pericare Wipes: SureStep" Foley Tray System Add-A-Foley Catheter REF: A153200 UDI code: 00801741121067 10801741121064 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A304400A UDI code: 00801741073847 10801741073844 SureStep" Foley Tray System Add-A-Foley Catheter REF: A399400 UDI code: 00801741100369 10801741100366 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A399400A UDI code: 00801741073915 10801741073912 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A904400A UDI code: 00801741074097 10801741074094 SureStep Intermittent Catheter Tray REF: INTAAI UDI code: 00801741183317 10801741183314 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.
Status
Open, Classified
Initiated
2026-06-25
Posted
2026-09-04
Terminated
not on file
Reason
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.
Root cause
Under Investigation by firm
Action
On 06/25/2026, an "URGENT Medical Device Recall" Letter was via regional courier services or other electronic means to inform customers that based on customer feedback, BD has identified that the CliniClean Pericare Wipes provided within Urinary Catheter Trays may be potentially contaminated with an unknown non-biological substance. Customer are instructed to: Do not use CliniClean Pericare Wipes contained within the affected products listed in Attachment 2. The CliniClean Pericare Wipes are the only impacted component within the tray and all remaining components can continue to be safely used. Users should source alternative, cleansing wipes for the impacted product lots listed in Attachment 2 and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters. If affected CliniClean Pericare Wipes have been used without issue, no additional treatment or clinical follow up is required. 1. For affected lots, remove and destroy wipes within the product package at the point of use and do not use. Source alternative cleansing wipes for all impacted urinary catheter trays listed in Attachment 2. 2. Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Correction. 3. Post this notice in all locations where inventory of these trays are stored. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may confirm your receipt of this notification. For Questions - contact: BD Medical Information Services: Available 24/7 Select #2 when prompted 1-800-555-7422 or Email: medical.information@bd.com BD Technical Support 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted M-F 8am - 5pm CT or Email: productcomplaints@bd.com
Quantity
33850 units/kits
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY. The countries of Bahrain, Canada, Costs Rica, Guatemala, Mexico, Saudi Arabia, and United Arab Emirates.
Lot, serial or model codes
SureStep" Foley Tray System Add-A-Foley Catheter A153200 / 00801741121067¿ 10801741121064 Lot Number: NGKS2397 SureStep" Foley Tray System Complete Care¿ Add-A-Foley Catheter A304400A / 00801741073847¿ 10801741073844 Lot Numbers: NGKU1023 NGKU3043 SureStep" Foley Tray System Add-A-Foley Catheter A399400 / 00801741100369¿ 10801741100366 Lot Number: NGKU2042 SureStep" Foley Tray System Complete Care¿ Add-A-Foley Catheter A399400A / 00801741073915¿ 10801741073912 Lot Numbers: NGKT2456 NGKT3292 NGKU0999 NGKU1026 SureStep" Foley Tray System Complete Care¿ Add-A-Foley Catheter A904400A / 00801741074097¿ 10801741074094 Lot Numbers: NGKT3231 NGKU2051 SureStep Intermittent Catheter Tray INTAAI / 00801741183317 10801741183314 Lot Number: NGKU1409
510(k) numbers
not on file
PMA numbers
not on file
Product code
KNX
Firm FEI
1018233
Firm city
Covington
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register