FDA Medical Device Recalls (openFDA)
Zimmer Biomet, Inc.: Herga foot switch, model 6289-WS, a component of the ROSA Robotized Stereotactic Assistant Surgical Device, Model 2.5.8. The firm name on the foot switch label is Herga Electric Limited, Bury, St. Edmunds, Suffolk IP32 6NN. ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
- Recall number
- Z-3117-2017
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- Herga foot switch, model 6289-WS, a component of the ROSA Robotized Stereotactic Assistant Surgical Device, Model 2.5.8. The firm name on the foot switch label is Herga Electric Limited, Bury, St. Edmunds, Suffolk IP32 6NN. ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
- Status
- Terminated
- Initiated
- 2016-01-20
- Posted
- not on file
- Terminated
- 2018-04-30
- Reason
- Complaints were received reporting the system would freeze/shut down while in Fulgurate mode.
- Root cause
- Software design
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Correction letter dated July 26, 2017, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product. for all surgical procedures, notify and ensure affected personnel are aware of the contents, complete the attached Certificated of Acknowledgement and return a digital copy. Customers with questions were instructed to call +33 (0) 467 107740, US customers should call 800-874-7711, ext 9225 .The initial correction was conducted by the manufacturer on January 20, 2016 with two of the 32 affected units which were imported into the U.S. being corrected by a customer representative. For questions regarding this recall call 574-267-6131.
- Quantity
- 64 devices
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Canada, China, France, Germany, India, Israel, Italy, Russia, Saudi Arabia, Spain and the United Kingdom.
- Lot, serial or model codes
- Serial numbers for ROSA devices with the Herga Footswitch in US distribution: Serial No. RO10009, RO10011, RO10014, RO13023, RO13027, RO14033, RO14035, RO14037, RO14038, RO14039, RO14040, RO14041, RO14043, RO15044, RO15045, RO15046, RO15048, RO15050, RO15051, RO15052, RO15053, RO15054, RO15058, RO15059, RO15060, RO15061, RO15062, RO15063, RO15064, RO15067, RO15069, and RO14031. Serial numbers for ROSA devices with the Herga Footswitch in foreign distribution: Serial No. RO08003, RO09004, RO09005, RO10007, RO10010, RO10013, RO11015, RO11016, RO11017, RO12018, RO12019, RO12020, RO12021, RO12022, RO13024, RO13025, RO13026, RO14029, RO14030, RO14032, RO14034, RO14036, RO14042, RO15047, RO15049, RO15055, RO15056, RO15057, RO15065, RO15066, RO15068, and RO15070.
- 510(k) numbers
- ["K092239","K101791"]
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28