Implant Direct Sybron Manufacturing, LLC: Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.
- Recall number
- Z-3115-2017
- Recalling firm
- Implant Direct Sybron Manufacturing, LLC
- Product
- Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.
- Status
- Terminated
- Initiated
- 2017-06-26
- Posted
- not on file
- Terminated
- 2021-06-16
- Reason
- The inside diameter of the 2.3mm small drill guide inserts have a smaller diameter than the specifications require.
- Root cause
- Process control
- Action
- Implant Direct sent an Safety Alert /Advisory Notice letter dated June 26, 2017. Instructions included to review and remove inventory for return to Implant Direct, complete and return the acknowledgement and recall return form, and contact customers and notify the of the recall if product was further distributed. For further questions, please call (888) 649-6425.
- Quantity
- 84 units
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution to the states of : AZ, CA, CO, CT, FL, ID, IL, KS, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA., and to the countries of : Canada, Netherlands, Spain, Great Britain, Germany and Switzerland.
- Lot, serial or model codes
- Lot 65249, Sequence Number HE G-HKRev01, Device Listing Number D245883.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3001617766
- Firm city
- Westlake Village
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28