Skip to the content

FDA Medical Device Recalls (openFDA)

Implant Direct Sybron Manufacturing, LLC: Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.

Recall number
Z-3115-2017
Recalling firm
Implant Direct Sybron Manufacturing, LLC
Product
Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.
Status
Terminated
Initiated
2017-06-26
Posted
not on file
Terminated
2021-06-16
Reason
The inside diameter of the 2.3mm small drill guide inserts have a smaller diameter than the specifications require.
Root cause
Process control
Action
Implant Direct sent an Safety Alert /Advisory Notice letter dated June 26, 2017. Instructions included to review and remove inventory for return to Implant Direct, complete and return the acknowledgement and recall return form, and contact customers and notify the of the recall if product was further distributed. For further questions, please call (888) 649-6425.
Quantity
84 units
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AZ, CA, CO, CT, FL, ID, IL, KS, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA., and to the countries of : Canada, Netherlands, Spain, Great Britain, Germany and Switzerland.
Lot, serial or model codes
Lot 65249, Sequence Number HE G-HKRev01, Device Listing Number D245883.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3001617766
Firm city
Westlake Village
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register