FDA Medical Device Recalls (openFDA)

Atrium Medical Corporation: Getinge PVC Thoracic Catheter, Firm PVC Straight, 10 per Case; Size(Model Number): 1. 28FR(15028), 2. 36FR(15036);

Recall number
Z-3113-2026
Recalling firm
Atrium Medical Corporation
Product
Getinge PVC Thoracic Catheter, Firm PVC Straight, 10 per Case; Size(Model Number): 1. 28FR(15028), 2. 36FR(15036);
Status
Open, Classified
Initiated
2026-08-07
Posted
2026-09-04
Terminated
not on file
Reason
An Out of Specification result (OOS) related to bacterial endotoxin testing in a single lot of catheters.
Root cause
Nonconforming Material/Component
Action
On August 7, 2026 URGENT MEDICAL DEVICE - RECALL letters were sent to customers. Customer Actions: 1. Please examine your inventory immediately to determine if you have any of the Atrium PVC Thoracic Catheters with the specific lot numbers. 2. Should you have any affected product, please remove from area(s) of use and quarantine. 3. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 4. Please contact Contact Customer Service via email at customersupport.us@getinge.com or 888-943-8872, option 2 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone) to request a return authorization (RMA) and shipping instructions to return any affected product. 5. Whether or not you have affected product(s) within expiration listed in this notice, please complete and sign the attached MEDICAL DEVICE- REMOVAL RESPONSE FORM to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.
Quantity
256
Distribution
US Nationwide distribution.
Lot, serial or model codes
1. Model Number(Size): 15028(28FR); UDI-DI: 00650862150285; Lot Number: ME244891; 3. Model Number(Size): 15036(36FR); UDI-DI: 00650862150360; Lot Number: ME244850;
510(k) numbers
not on file
PMA numbers
not on file
Product code
GBZ
Firm FEI
3011175548
Firm city
Merrimack
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register