Toshiba American Medical Systems: Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with the ceiling suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and trans mission/ output of images to external devices.
- Recall number
- Z-3111-2017
- Recalling firm
- Toshiba American Medical Systems
- Product
- Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with the ceiling suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and trans mission/ output of images to external devices.
- Status
- Terminated
- Initiated
- 2017-03-23
- Posted
- not on file
- Terminated
- 2020-10-23
- Reason
- It was discovered during a procedure that when the operator made an exposure on the wireless x-ray detector and the image data was sent to the digital radiography system, and error message was displayed " System Error (2063)" which required a reboot and loss of the image.
- Root cause
- Software design
- Action
- Toshiba America Medical Systems Inc. planned action to bring defect into compliance: 1.You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2.You will install modified firmware on the systems to correct the defect and prevent recurrence. 3.You will provide progress reports on the effectiveness of the plan. If you have any questions, please feel free to contact your local Toshiba representative at (800) 421-1968.
- Quantity
- 140
- Distribution
- USA (nationwide) Distribution
- Lot, serial or model codes
- Serial No. W8B1572022
- 510(k) numbers
- ["K122842"]
- PMA numbers
- not on file
- Product code
- MQB
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28