Philips Healthcare Inc.: Philips Medical: 865047 IntelliVue Clinical Information Portfolio Release D.03 865209 IntelliVue Clinical Information Portfolio Release E.00 and E.01 865300 IntelliVue Clinical Information Portfolio NICU Link Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.
- Recall number
- Z-3111-2011
- Recalling firm
- Philips Healthcare Inc.
- Product
- Philips Medical: 865047 IntelliVue Clinical Information Portfolio Release D.03 865209 IntelliVue Clinical Information Portfolio Release E.00 and E.01 865300 IntelliVue Clinical Information Portfolio NICU Link Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.
- Status
- Terminated
- Initiated
- 2011-07-08
- Posted
- 2011-08-31
- Terminated
- 2020-08-13
- Reason
- Under certain circumstances, edits to an order can result in unintended scheduled interventions on the IntelliVue Clinical Information Portfolio. A patient may receive interventions or additional medication doses that are not intended.
- Root cause
- Software design
- Action
- Philips Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated July 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers are asked to follow the guidelines in the "Action to be taken by Customer/User" section of the Urgent Medical Device Correction Notice. Philips has developed a software upgrade to address this problem in the short term. This upgrade will be available immediately. This software upgrade will be performed by Philips service personnel. While customers await the correction of their devices, customers are instructed to follow the instructions for use for the device, which includes the directions to: 1. Always use the medication administration record (MAR) in determining patient care. 2. Always review orders and pending administrations in the administration record after any changes to an order are stored. For the longer term, Philips is developing safeguards intended to minimize the potential for similar errors resulting from order changes and upgrade to be available in September 2011. If you need any further information or support concerning this issue, please contact your local Philips representative.
- Quantity
- 454 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY HONG KONG, IRELAND, ITALY, KENYA, LATVIA, MALAYSIA, NETHERLANDS, NORWAY, SAUDI ARABIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, UNITED ARAB EMIRATES, and UNITED KINGDOM.
- Lot, serial or model codes
- Software version: D.03, E.0, or E.01 of the software
- 510(k) numbers
- ["K100272","K992636"]
- PMA numbers
- not on file
- Product code
- DXJ
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28