Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Inc.: Philips Medical: 865047 IntelliVue Clinical Information Portfolio Release D.03 865209 IntelliVue Clinical Information Portfolio Release E.00 and E.01 865300 IntelliVue Clinical Information Portfolio NICU Link Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.

Recall number
Z-3111-2011
Recalling firm
Philips Healthcare Inc.
Product
Philips Medical: 865047 IntelliVue Clinical Information Portfolio Release D.03 865209 IntelliVue Clinical Information Portfolio Release E.00 and E.01 865300 IntelliVue Clinical Information Portfolio NICU Link Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.
Status
Terminated
Initiated
2011-07-08
Posted
2011-08-31
Terminated
2020-08-13
Reason
Under certain circumstances, edits to an order can result in unintended scheduled interventions on the IntelliVue Clinical Information Portfolio. A patient may receive interventions or additional medication doses that are not intended.
Root cause
Software design
Action
Philips Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated July 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers are asked to follow the guidelines in the "Action to be taken by Customer/User" section of the Urgent Medical Device Correction Notice. Philips has developed a software upgrade to address this problem in the short term. This upgrade will be available immediately. This software upgrade will be performed by Philips service personnel. While customers await the correction of their devices, customers are instructed to follow the instructions for use for the device, which includes the directions to: 1. Always use the medication administration record (MAR) in determining patient care. 2. Always review orders and pending administrations in the administration record after any changes to an order are stored. For the longer term, Philips is developing safeguards intended to minimize the potential for similar errors resulting from order changes and upgrade to be available in September 2011. If you need any further information or support concerning this issue, please contact your local Philips representative.
Quantity
454 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY HONG KONG, IRELAND, ITALY, KENYA, LATVIA, MALAYSIA, NETHERLANDS, NORWAY, SAUDI ARABIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, UNITED ARAB EMIRATES, and UNITED KINGDOM.
Lot, serial or model codes
Software version: D.03, E.0, or E.01 of the software
510(k) numbers
["K100272","K992636"]
PMA numbers
not on file
Product code
DXJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register