Elekta, Inc.: MOSAIQ Medical Charged-Particle Radiation Therapy System
- Recall number
- Z-3110-2017
- Recalling firm
- Elekta, Inc.
- Product
- MOSAIQ Medical Charged-Particle Radiation Therapy System
- Status
- Terminated
- Initiated
- 2017-06-30
- Posted
- not on file
- Terminated
- 2020-10-26
- Reason
- The table shift in the vertical direction was inverted during the use of CMA (Couch Move Assistant). Therapy may be directed to an incorrect area.
- Root cause
- Software in the Use Environment
- Action
- Consignees were to receive a recall notice by mail dated 06/30/2017. The notice identifies the reason for recall and goes on to instruct consignees to run verification tests on units to ensure they are operating properly. Only applicable tests should be performed on the unit. If the unit passes verification tests, then the attached acknowledgment form is to be signed and returned to Elekta. If the unit fails verification, consignees are instructed to contact their local Elekta Care Support Center for assistance. If a unit is no longer in use, consignees are asked to sign and return the acknowledgement form, noting that the unit is no longer in use. The firm asks that the recall notice be placed in an accessible area to users until the field action is closed. Appropriate personnel should also be provided the content of the recall notification.
- Quantity
- 2,461
- Distribution
- Worldwide distribution. US nationwide to AK, AL, AR, AZ, CA, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, District of Columbia, Hawaii and Puerto Rico. ALBANIA, ALGERIA, ANTIGUA AND BARBUDA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BENIN, BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CUBA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GABON, GEORGIA, GERMANY, GHANA, GREECE, HONDURRAS, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, LAOS, LIBYA, LITHUANIA, MALAYSIA, MALTA, MEXICO, MOROCCO, MYANMAR, NAMIBIA, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, VENEZUELA, VIETNAM and ZIMBABWE
- Lot, serial or model codes
- All software builds.
- 510(k) numbers
- ["K141572"]
- PMA numbers
- not on file
- Product code
- LHN
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28