Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: REF 91200, Juggerknot system, Soft Anchors, -Two implants size 1, Biomet Sports Medicine, STERILE. Product Usage: The JuggerKnot Soft Anchors include various size ranges, each consisting of a coreless sleeve and suture(s) assembled on an inserter instrument. The system includes instrumentation allowing for proper preparation and placement of the soft tissue fixation device.

Recall number
Z-3109-2011
Recalling firm
Biomet, Inc.
Product
REF 91200, Juggerknot system, Soft Anchors, -Two implants size 1, Biomet Sports Medicine, STERILE. Product Usage: The JuggerKnot Soft Anchors include various size ranges, each consisting of a coreless sleeve and suture(s) assembled on an inserter instrument. The system includes instrumentation allowing for proper preparation and placement of the soft tissue fixation device.
Status
Terminated
Initiated
2011-07-21
Posted
2011-08-31
Terminated
2013-04-26
Reason
The device packaging was punctured by the device, resulting in loss of implant sterility.
Root cause
Package design/selection
Action
Biomet Sports Medicine notified distributor accounts by telephone on July 08, 2011 with instructions to quarantine all lots of the recalled product. Urgent Medical Device Recall Notice letters dated July 21, 2011 were sent to all affected customers. The letter identified the recalled product, problem, and actions to be taken. The customers were instructed to immediately locate and discontinue use of the recalled product. The distributors are advised to communicate this notice to their customers if the recalled products were further distributed. The letter also states that distributors are responsible for the return of the recalled product. Customers were asked to complete a response form and fax to 574-372-1683, within 3 business days. Customers were instructed to provide phone confirmation at 800-348-9500, extension 3755 or 3756, when the recall notice is received. Further questions should be directed to 574-371-3755 or 574-371-3756, Monday through Friday, 8 AM to 5 PM.
Quantity
4925
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Columbia, Europe and Korea.
Lot, serial or model codes
Lot #: M050020, Ml02470, M205970, M206170, M335840, M451670, M554290, M592790, M906510, M050030, M102480, M205980, M206180, M335860, M451680, M554300, M596210, M906540, M050040, M102490, M205990, M206200, M335870, M451690, M554310, M666350, M906550, M050050, M102500, M206000, M213400, M336170, M451700, M555140, M666360, M974690, M050060, M102510, M206010, M213410, M350540, M475050, M571420, M725330, M974720, M050070, M102520, M206030, M213420, M350550, M479960, M571560, M725340, M050080, M102530, M206060, M248050, M350560, M506760, M571570, M727850, M051940, M102540, M206070, M248060, M350570, M513940, M571580, M762160, M051950, M185150, M206080, M248070, M351000, M537410, M571600, M802400, M051960, M185160, M206090, M248080, M354670, M554260, M571730, M802410, M051970, M205940, M206140, M248090, M451650, M554270, M571760, M802420, M072920, M205960, M206150, M248100, M451660, M554280, M571800, and M860160
510(k) numbers
["K110145"]
PMA numbers
not on file
Product code
MBI
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register