Skip to the content

FDA Medical Device Recalls (openFDA)

Spectranetics Corporation: Spectranetics Bridge Occlusion Balloon catheter, Sheath Compatibility 12F, Guidewire Compatibility 0.035in, Working Length 90cm, Balloon Outer Diameter 31mm, Balloon Length 80mm, Rx Only. Product Usage: It is a device used to occlude the vessel and prevent blood loss as the physician preps for surgery resulting from an SVC tear complication of a lead extraction case. It is used for temporary vessel occlusion of the superior vena cava in applications including perioperative occlusion and emergency control of hemorrhage.

Recall number
Z-3106-2017
Recalling firm
Spectranetics Corporation
Product
Spectranetics Bridge Occlusion Balloon catheter, Sheath Compatibility 12F, Guidewire Compatibility 0.035in, Working Length 90cm, Balloon Outer Diameter 31mm, Balloon Length 80mm, Rx Only. Product Usage: It is a device used to occlude the vessel and prevent blood loss as the physician preps for surgery resulting from an SVC tear complication of a lead extraction case. It is used for temporary vessel occlusion of the superior vena cava in applications including perioperative occlusion and emergency control of hemorrhage.
Status
Terminated
Initiated
2017-08-07
Posted
2017-08-26
Terminated
2020-04-14
Reason
Possible inability to pass the guidewire through the Bridge device, therefore, resulting in a potential delay of treatment.
Root cause
Nonconforming Material/Component
Action
The firm, Spectranetics, sent an "MEDICAL DEVICE RECALL" letter dated 08/07/2017 to customers via certified mail and an updated letter dated 8/17/2017 was sent on 8/23/2017. The letter described the product, problem and actions to be taken. The customers were Instructed to do the following: 1. For all Bridge inventory, we recommend physicians continue to follow product instructions to place a guide wire through the venous access site prior to the start of the procedure. Physicians may consider removing Bridge from packaging and placing the Bridge balloon over the wire prior to the start of the procedure to ensure a patent guidewire lumen if they believe the case is high risk. 2. All Bridge inventory will continue to have the potential for a guidewire lumen blockage until a solution is implemented. 3. Please bring your inventory into the case so that you have the back-up on hand, should you need it. 4. Always follow appropriate complication prevention and management protocols as it relates to patient preparation and surgical back up for lead extraction procedures. 5. Please complete the attached Acknowledgement and Receipt Form by EMAIL: Customer.Service@SPNC.com or FAX to: 1-877-447-2022. Your Spectranetics Sales Representative will be contacting you to facilitate the return and replacement of any remaining product in inventory once new inventory becomes available with an implemented fix; however, you may also reach Customer Service at 1-800-231-0978, Option 2. If you have additional questions please feel free to discuss with your local Spectranetics Sales Representative, or call the Vice President Quality Assurance directly at 719-447-2469. The Spectranetics Customer Service Department is also available to support you with any assistance you may need. Contact information:1-800-231-0978, Option 2 (Hours of Operation- Monday- Friday 7:00AM 5:00PM MST) or Email: Customer.service@spnc.com. An additional letter was sent via electronic notific
Quantity
2,356 units
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Cyprus, Denmark, Finland, France, Germany, Hungary, Iceland, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Poland, South Africa, Sweden, Switzerland, and United Kingdom.
Lot, serial or model codes
Model Number 590-001, Lots FMN17B13A, FMN17C08A, FMN17C28A, FMN17D07A, FMN47D12A, FMN17D19A, FMN17D27A, FMN17E02A, FMN17E23A, FMN17E31A, FMN17E31B, FMN17F06A, FMN17F20A, FMN17F21A, FMN17G12A, FMN17G18A.    Additional lots: FMN17H03A, FMN17H04A, FMN17H05A, FMN17H10A, FMN17H12A, FMN17H15A, FMN17H17A, FMN17H29A, FMN17J28A, FMN17K02A.
510(k) numbers
["K153530"]
PMA numbers
not on file
Product code
MJN
Firm FEI
3007284006
Firm city
Colorado Springs
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register