Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Mitek, Inc., a Johnson & Johnson Co.: Depuy Mitek Ratchet Handle with Quick Connect Product code: 219970. Instrument used with anchor inserters for insertion of DePuy Mitek Interference Screws (orthopedic).

Recall number
Z-3106-2011
Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
Product
Depuy Mitek Ratchet Handle with Quick Connect Product code: 219970. Instrument used with anchor inserters for insertion of DePuy Mitek Interference Screws (orthopedic).
Status
Terminated
Initiated
2011-06-27
Posted
2011-08-31
Terminated
2012-01-19
Reason
Quick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head.
Root cause
Process change control
Action
Depuy Mitek issued Original Customer Notification Letter on 6/27/11 and an Updated letter to clarify the lot number information sent 6/29/11. Hospitals requested to immediately check all inventories for affected product and return for repalcement product. Also notify surgeons aware of this potential issue and potential risk of a delay in surgery. questions or concerns with regard to this program, please contact your local DePuy Mitek Representative or DePuy Mitek Customer Service directly at 1-800-382-4682, PROMPT 3.
Quantity
75 units
Distribution
Nationwide distribution.
Lot, serial or model codes
Finished Goods (Box label) Lot Numbers: 1105101 and 1106132. Component lot number Etched on Reusable Handle 1012001 and 1106001
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1221934
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register