Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems

Recall number
Z-3105-2017
Recalling firm
Merge Healthcare, Inc.
Product
Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems
Status
Terminated
Initiated
2017-08-14
Posted
2017-08-18
Terminated
2019-05-08
Reason
The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
Root cause
Process design
Action
MERGE sent a Medical Device Recall letter dated August 11, 2017, to all affected customers. Customers were advised to discontinue using the affected product and to reply using the enclosed form. Customers with questions were instructed to send an email to recall@merge.com. For questions regarding this recall call 262-912-3514.
Quantity
98 sites have the potentially affected software
Distribution
Nationwide Distribution
Lot, serial or model codes
Software versions 10.0, 10.0.1, 10.0.2, 10.0.3, 10.0.4, 10.0.5, 11.0, 11.0.1, 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1, 11.1.1.1, 11.1.1 Patch 2, 11.1.1 Patch 3, 11.1.2, 11.1.2 Patch 1, 11.1.2 Patch 2, 11.1.2 Patch 3, and 11.1.2 Patch 4.  Only customers with one of the above listed software versions AND a license for Mammo Tracking are impacted by this recall.
510(k) numbers
["K041935","K143318"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register