Skip to the content

FDA Medical Device Recalls (openFDA)

Stanmore Implants Worldwide Ltd.: JTS External Drive Unit JTS Distal Femoral Extendible Implant Femur is intended for the replacement of diseased or deficient bone in the distal femur.

Recall number
Z-3104-2017
Recalling firm
Stanmore Implants Worldwide Ltd.
Product
JTS External Drive Unit JTS Distal Femoral Extendible Implant Femur is intended for the replacement of diseased or deficient bone in the distal femur.
Status
Terminated
Initiated
2017-07-04
Posted
not on file
Terminated
2021-02-16
Reason
Three (3) complaints were identified in which the potential for utilization of the JTS Grower Drive in the incorrect orientation in relation to the limb entry point of the device, together with the potential to select incorrect settings for the extension procedure. As a result, the prostheses in such cases failed to extend as intended.
Root cause
Labeling design
Action
Stanmore Implants sent an Urgent Field Safety Notice dated July 4, 2017, notifying customers of the recall. Additionally, the recalling firm requests that consignees immediately check their inventory for recalled product, quarantine any product on hand, inform SIW of any product on hand, complete the attached response form, and return the response for and affected product to the firm. It is requested that consignees distribute the notice internally to all affected parties and that they contact their local Stanmore representative to request training. If any adverse events are associated with this device, the firm asks that it be reported to them. For further questions, please call (972) 501-1402.
Quantity
15 units
Distribution
US Distribution to the states of : CA, FL, GA, KS, MI, NJ, OH, PA, SC, and TX.
Lot, serial or model codes
Product Code MLE3; Serial No. 907-103, 907-104, 907-119, 907-123, 907-125, 907-126, 907-127, 907-028, 907-237, 907-240, 907-500, 907-501, 907-502, 907-503, 907-504.
510(k) numbers
["K092138","K133152","K140898"]
PMA numbers
not on file
Product code
KRO
Firm FEI
3004105610
Firm city
Borehamwood
Firm state
Centennial Park
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register