Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, Lp: MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58

Recall number
Z-3102-2017
Recalling firm
Encore Medical, Lp
Product
MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
Status
Terminated
Initiated
2017-08-04
Posted
not on file
Terminated
2023-06-28
Reason
It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
Root cause
Employee error
Action
The firm initiated their recall by telephone and email to their sales representatives and surgeons on 08/04/2017. The letter explained the problem and advised the surgeons that the sales representative would be collecting the devices for return.
Quantity
5 units
Distribution
OH, VA
Lot, serial or model codes
Catalog number, Lot Number S-200956, 535U1000 S-200957, 536U1000 S-200958, 537U1000
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTO
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register