FDA Medical Device Recalls (openFDA)
Encore Medical, Lp: MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
- Recall number
- Z-3102-2017
- Recalling firm
- Encore Medical, Lp
- Product
- MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
- Status
- Terminated
- Initiated
- 2017-08-04
- Posted
- not on file
- Terminated
- 2023-06-28
- Reason
- It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
- Root cause
- Employee error
- Action
- The firm initiated their recall by telephone and email to their sales representatives and surgeons on 08/04/2017. The letter explained the problem and advised the surgeons that the sales representative would be collecting the devices for return.
- Quantity
- 5 units
- Distribution
- OH, VA
- Lot, serial or model codes
- Catalog number, Lot Number S-200956, 535U1000 S-200957, 536U1000 S-200958, 537U1000
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HTO
- Firm FEI
- 1000116912
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as