Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System, Software Version 3.2. The DxH Slidemaker Stainer allows for adaptation of the smear appearance and stain methodology according to user preferences. Blood smears produced by the Slidemaker portion of the DxH Slidemaker Stainer are moved to baskets for transfer to the Stainer portion by a robot arm. Microscopic examination of the stained blood smears can be used to help determine the hematologic status of a patient. A stained blood film: Allows for the differentiation of white blood cells, facilitates the characterization of red blood cells and platelets, and aids in the identification of blood components and cellular abnormalities.

Recall number
Z-3101-2017
Recalling firm
Beckman Coulter Inc.
Product
UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System, Software Version 3.2. The DxH Slidemaker Stainer allows for adaptation of the smear appearance and stain methodology according to user preferences. Blood smears produced by the Slidemaker portion of the DxH Slidemaker Stainer are moved to baskets for transfer to the Stainer portion by a robot arm. Microscopic examination of the stained blood smears can be used to help determine the hematologic status of a patient. A stained blood film: Allows for the differentiation of white blood cells, facilitates the characterization of red blood cells and platelets, and aids in the identification of blood components and cellular abnormalities.
Status
Terminated
Initiated
2017-06-20
Posted
not on file
Terminated
2021-04-21
Reason
The firm has identified that a fitting in some Slidemaker Stainer units may not completely seal after installation.
Root cause
Employee error
Action
Beckman Coulter sent an Urgent Medical Device Recall letter dated June 20, 2017. The notification advised consignees to immediately contact a Beckman Coulter representative if a leak is observed to schedule a service, to inspect reagent supply drawer for spills, and to follow best laboratory practices when using the device (including Personal Protective Equipment). Beckman Coulter indicates that the issue will be resolved with the replacement of the correct fitting on all impacted instruments. Labs were advised to notify all staff of the recall and to retain the notification. They were also instructed to forward the recall notice to all customers, if any product was further distributed, and to complete and return the Customer Response form within ten days of receiving the recall notification. Completed forms are to be sent back to the firm via email to Regulatory.Notifications@beckmancoulter.com. Regulatory Affairs Mail Code 31-B06. Questions can be directed to beckmancoulter.com or 800-526-7694. Consignees outside of the US are advised to contact their Beckman Coulter representative in the case that questions do arise.
Quantity
310 units
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AR, AZ, CA, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, and WV., and to the countries of : Australia, Brazil, Canada, Chile, China, Czech Republic, Denmark, France, Greece, Hong Kong, India, Israel, Italy, Japan, Morocco, Portugal, South Korea, South Africa, Spain, Taiwan, Turkey, United Kingdom, Vietnam, and Switzerland.
Lot, serial or model codes
P/N A85371AA20; Catalog No. 775222
510(k) numbers
["K140911"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register