Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: smith&nephew POLARSTEM(TM) Femoral Stems with Ti/HA (Standard, Lateral, Valgus and Collar), REF 76102260, NON STERILE, QTY: (1), Rx only The subject cardan is used in conjunction with a curved stem impactor used to implant the POLARSTEM hip stem. The hip stem is mounted on the screw at the tip of the instrument, and the stem is then impacted into the femur. After impaction, the surgeon uses a knob at the end of the cardan to unscrew the implanted stem from the instrument

Recall number
Z-3100-2017
Recalling firm
Smith & Nephew, Inc.
Product
smith&nephew POLARSTEM(TM) Femoral Stems with Ti/HA (Standard, Lateral, Valgus and Collar), REF 76102260, NON STERILE, QTY: (1), Rx only The subject cardan is used in conjunction with a curved stem impactor used to implant the POLARSTEM hip stem. The hip stem is mounted on the screw at the tip of the instrument, and the stem is then impacted into the femur. After impaction, the surgeon uses a knob at the end of the cardan to unscrew the implanted stem from the instrument
Status
Terminated
Initiated
2017-07-24
Posted
not on file
Terminated
2020-08-21
Reason
The Cardan joint of the Polarstem Cardan could potentially fracture or functionally fail after multiple procedures.
Root cause
Device Design
Action
Smith & Nephew sent an Urgent Medical Device Recall Notice dated July 24, 2017, to all affected consignees. The firm notified their consignees by email and Fed Ex on 07/24/2017. Consignees have been asked to inspect inventory to locate and quarantine affected product. All consignees were asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew. Customers with questions were instructed to contact FieldActions@smith-nephew.com. For questions regarding this recall call 978-749-1440.
Quantity
402 units
Distribution
Worldwide Distribution - US (Nationwide) and Internationally to Australia, Belgium, Norway, and Switzerland.
Lot, serial or model codes
Lot Numbers: A57835, A57837, A58509, A59084, A59446, A59547, A60948
510(k) numbers
["K130728"]
PMA numbers
not on file
Product code
LZO
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register