Smith & Nephew Inc.: ROTATION MEDICAL TENDON STAPLES (8) (narrow indication)-Intended for fixation of prosthetic material to soft tissues. Product Number: 72205201

Recall number
Z-3097-2024
Recalling firm
Smith & Nephew Inc.
Product
ROTATION MEDICAL TENDON STAPLES (8) (narrow indication)-Intended for fixation of prosthetic material to soft tissues. Product Number: 72205201
Status
Open, Classified
Initiated
2024-08-12
Posted
2024-09-10
Terminated
not on file
Reason
Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility
Root cause
Packaging
Action
Smith & Nephew issued Urgent Medical Device Recall Notice R-2024-06 on 8/12/24. Letter states reason for recall, health risk and action to take: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customer s account number and name in Section A. 2. If you have no product to return: a. Please put an X in the appropriate box below. b. Complete Section B of the form, sign and send to FieldActions@smith-nephew.com. 3. If you have product to return: a. Please complete Section A - list the product part number, batches, and quantities that you are returning. b. Complete Section B of the form, sign and send to FieldActions@smith-nephew.com. c. Once the form is received by Smith+Nephew, you will be sent a Return Authorization (RA) number. d. Return affected product to: Smith+Nephew | Attn: Quality Hold and Field Action Returns Department | 76 South Meridian Avenue | Oklahoma City, OK 73107 If you have any questions or concerns regarding this recall please contact FieldActions@smithnephew.com
Quantity
2961 units (OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Chile, China, France, Germany, Great Britain, Greece, Hong Kong, India, Italy, Japan, Lithuania, Malaysia, Malta, New Zealand, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovenia, South Africa, Spain, Switzerland, United Arab Emirates/Dubai (UAE).
Lot, serial or model codes
UDI-DI: 00885556733486 Batch Numbers: 51192136 51192143 51192144 51192145 51192146 51192148 51192150 51192151 51192152 51192153 51192154 51204480 51204481 51204484 51204485 51204486 51204487 51204488 51204489 51204491 51223193 51223241 51223242 51223243 51223249 51224282
510(k) numbers
["K131637"]
PMA numbers
not on file
Product code
GDW
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register