Wipro GE Healthcare Private Ltd.: GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.

Recall number
Z-3081-2024
Recalling firm
Wipro GE Healthcare Private Ltd.
Product
GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
Status
Open, Classified
Initiated
2024-08-09
Posted
2024-09-20
Terminated
not on file
Reason
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
Root cause
Process control
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/09/2024 by letter disseminated through a traceable means. The notice explained the safety issue and the potential for adverse impact to user or patient. The notice also provided information for the safe operation of the device until GE Healthcare is able to inspect and test the units. A GE Healthcare representative will contact the consignee to arrange for the correction. Questions: Contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity
137 units
Distribution
US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.
Lot, serial or model codes
1) M1112198-01045518, UDI/DI 00840682152709, Serial Numbers: PBW23420278SA, PBW23390175SA, PBW24241071SA, PBW24241072SA, PBW24241075SA, PBW23340090SA; 2) M1112198-01045579, UDI/DI 00840682152709, Serial Numbers: PBW24241078SA, PBW24241079SA, PBW24241063SA, PBW24241076SA; 3) M1112198-01047148, UDI/DI 00840682152709, Serial Numbers: PBW24241068SA, PBW24241070SA, PBW24241065SA, PBW24241064SA, PBW24241066SA, PBW24241067SA, PBW24241069SA; 4) M1112198-01047271, UDI/DI 00840682152716, Serial Numbers: PBW23492144SA, PBW23482075SA, PBW23492149SA, PBW24062087SA, PBW24062078SA, PBW24062081SA, PBW23492145SA, PBW24152231SA, PBW24152225SA, PBW24152229SA, PBW23482085SA, PBW24202294SA, PBW24202292SA, PBW24202293SA, PBW24212323SA, PBW24212324SA, PBW24152226SA, PBW24152227SA, PBW24052060SA, PBW23492142SA; 5) M1112198-01048731, UDI/DI 00840682152709, Serial Numbers: PBW24241087SA, PBW24241086SA, PBW24241085SA, PBW23320042SA, PBW23320041SA, PBW24241084SA, PBW23320035SA; 6) M1112198-01048740, UDI/DI 00840682152716, Serial Numbers: PBW23482093SA; 7) M1112198-01052386, UDI/DI 00840682152716, Serial Numbers: PBW23482100SA, PBW24202309SA, PBW24202313SA, PBW24202310SA, PBW24202315SA, PBW24202312SA, PBW24202314SA, PBW24202311SA; 8) M1112198-01054697, UDI/DI 00840682152716, Serial Numbers: PBW24132200SA; 9) M1112198-01068942, UDI/DI 00840682152716, Serial Numbers: PBW23472056SA, PBW24062069SA, PBW24112144SA, PBW24062072SA, PBW24112157SA, PBW24112149SA, PBW24112145SA, PBW24112158SA, PBW24112148SA, PBW24112147SA, PBW24112155SA, PBW24112150SA, PBW24112153SA, PBW24112156SA, PBW24112152SA, PBW24112154SA, PBW24112146SA, PBW24112151SA, PBW24142209SA; 10) M1112198-01084972, UDI/DI 00840682152716, Serial Numbers: PBW23462025SA, PBW23462038SA, PBW23462039SA, PBW23472051SA, PBW23472049SA, PBW23472050SA, PBW24132204SA, PBW24132203SA, PBW24132205SA, PBW24122189SA, PBW24122190SA; 11) M1112198-01090848, UDI/DI 00840682152716, Serial Numbers: PBW24122198SA, PBW24122195SA, PBW24122197SA; 12) M1112198-01093457, UDI/DI 00840682152716, Serial Numbers: PBW24202308SA, PBW24132201SA; 13) M1112198-01117766, UDI/DI 00840682152709, Serial Numbers: PBW24110390SA; 14) M1112198-01139893, UDI/DI 00840682152716, Serial Numbers: PBW23242002SA, PBW23242003SA, PBW23242000SA, PBW23242001SA, PBW23512155SA, PBW24062070SA, PBW24172255SA, PBW24172253SA, PBW24172254SA, PBW24172251SA, PBW24172250SA, PBW24212327SA, PBW24212328SA; 15) M1112198-01156745, UDI/DI 00840682152716, Serial Numbers: PBW24192258SA; 16) M1112198-01167230, UDI/DI 00840682152716, Serial Numbers: PBW23462035SA, PBW23462036SA, PBW23462040SA, PBW24112159SA; 17) M1112198-01177090, UDI/DI 00840682152716, Serial Numbers: PBW23482099SA; 18) M1112198-01183599, UDI/DI 00840682152716, Serial Numbers: PBW23492140SA, PBW23482090SA, PBW23482092SA, PBW23482089SA, PBW23482086SA, PBW23492139SA, PBW23482087SA, PBW23482088SA, PBW23482091SA, PBW23482103SA, PBW23482104SA, PBW23482105SA; 19) M1112198-01185372, UDI/DI 00840682152716, Serial Numbers: PBW23482079SA, PBW23482081SA, PBW23482082SA, PBW23482083SA, PBW23482078SA; 20) M1112198-01193801, UDI/DI 00840682152716, Serial Numbers: PBW23502154SA, PBW24172252SA; 21) M1112198-01200764, UDI/DI 00840682152716, Serial Numbers: PBW23472043SA; 22) M1112198-01245548, UDI/DI 00840682152716, Serial Numbers: PBW24062101SA; 23) M1112198-01253742, UDI/DI 00840682152716, Serial Numbers: PBW24052062SA, PBW24052063SA, PBW24052066SA, PBW24052065SA, PBW24052064SA; 24) M1112198-01274059, UDI/DI 00840682152716, Serial Numbers: PBW24192278SA, PBW24192279SA.
510(k) numbers
["K201628"]
PMA numbers
not on file
Product code
FMT
Firm FEI
3007734888
Firm city
Bengaluru
Firm state
not on file
Firm country
India

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register