Skip to the content

FDA Medical Device Recalls (openFDA)

Orthohelix Surgical Designs Inc: Mini MaxLock Extreme Plating System Case & Tray, Model Number(s): MXM-901, Product Usage: The Mini MaxLock Extreme Plating System is intended to stabilize and aid in the repair of fractures, fusions, and osteotomies for small bones and bone fragments. Surgical Case and Tray Delivery System used to house plating and screw orthopedic implant system and instrumentation.

Recall number
Z-3081-2011
Recalling firm
Orthohelix Surgical Designs Inc
Product
Mini MaxLock Extreme Plating System Case & Tray, Model Number(s): MXM-901, Product Usage: The Mini MaxLock Extreme Plating System is intended to stabilize and aid in the repair of fractures, fusions, and osteotomies for small bones and bone fragments. Surgical Case and Tray Delivery System used to house plating and screw orthopedic implant system and instrumentation.
Status
Terminated
Initiated
2011-07-06
Posted
2011-08-29
Terminated
2012-09-10
Reason
The surgical tray does not meet our Instruction for Use (IFU) for dry time requirement.
Root cause
Nonconforming Material/Component
Action
Orthohelix Surgical Designs Inc sent a RECALL NOTIFICATION letter dated July 6, 2011 and an UPDATED URGENT RECALL 2ND NOTIFICATION letter dated July 13, 2011 to all affected customers. The letter identified the recalled product, the problem, and the Field Correction actions to be taken. Customers were instructed to remove the blue rubber mat in the utility bay of the tray base immediately, tag the mat with the tray number, and return it to recalling firm immediately. The letter states once customers have removed the blue rubber mat from the tray base, they may continue to use the tray and system normally. No further actions or precautions are needed. Distributors were also asked to notify their customers of the recall. If you have any questions, please call 330-247-1444 or 330-247-1445.
Quantity
61 trays
Distribution
USA Nationwide Distribution including the states of: AR, AZ, CA, CO, FL, GA, IL, IN, KY, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI & WY.
Lot, serial or model codes
Model Number(s): MXM-901; Tray codes affected: LAQ, LAV, LCL, LCN, LBZ, LCA, LBV, LCU, LAR, LBJ, LBT, LDE, LDG, LBL, LCD, LBI, LCI, LBQ, LCV, LBK, LBC, LCW, LCB, LDH, LBM, LCM, LCK, LDB, LBS, LCE, LDD, LDF, LAX, LDI, LAP, LBH, LBP, LAO, LBB, LBE, LBX, LCH, LAT, LAS, LCR, LDC, LAW, LCJ, LDJ, LCZ, LAU, LBF, LCQ, LBN, LCP, LCX, LBU, LBW, LBY, LCT, LCY, LAY, LAZ, LBA, LBD, LBG, LBO, LBR, LCC, LCF, LCG, LCS, LDA, LDM, LDN, LDO, LDP, LDQ, LDR, LDS, LDT, LDU & LDK.
510(k) numbers
["K101962"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3000718467
Firm city
Medina
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register