Skip to the content

FDA Medical Device Recalls (openFDA)

Sol-Millennium Medical Inc.: Syringes with Low Dead Space labeled as: a) SOL-M 1ml Slip Tip Syringe w/o needle (low dead space), Model No. 180011LDS, UPC 840368511899; b) SOL-M 1ml Slip Tip Syringe w/o needle (low dead space), Model No. P180011LDS, UPC 810018097554; c) SOL-M 1ml Luer Lock Syringe w/Exch Needle 23G*1" (low dead space), Model No. 181023L, UPC Sold in Canada Only; d) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 25G*5/8" low dead space), Model No. 12558LDS, UPC 810062435319; e) SOL-M 1ml Luer Lock Syringe w/o Needle (PP) (low dead space), Model No. P180001PP, UPC 818392018165; f) SOL-M 1ml TB Slip Tip Syringe w/Exch Needle 25G*5/8'' (low dead space), Model No. 181025S, UPC Sold in Canada Only

Recall number
Z-3078-2024
Recalling firm
Sol-Millennium Medical Inc.
Product
Syringes with Low Dead Space labeled as: a) SOL-M 1ml Slip Tip Syringe w/o needle (low dead space), Model No. 180011LDS, UPC 840368511899; b) SOL-M 1ml Slip Tip Syringe w/o needle (low dead space), Model No. P180011LDS, UPC 810018097554; c) SOL-M 1ml Luer Lock Syringe w/Exch Needle 23G*1" (low dead space), Model No. 181023L, UPC Sold in Canada Only; d) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 25G*5/8" low dead space), Model No. 12558LDS, UPC 810062435319; e) SOL-M 1ml Luer Lock Syringe w/o Needle (PP) (low dead space), Model No. P180001PP, UPC 818392018165; f) SOL-M 1ml TB Slip Tip Syringe w/Exch Needle 25G*5/8'' (low dead space), Model No. 181025S, UPC Sold in Canada Only
Status
Open, Classified
Initiated
2024-05-09
Posted
2024-09-06
Terminated
not on file
Reason
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
Root cause
Other
Action
Sol M issued a Voluntary Recall Action Letter to Distributors to its consignees on 05/09/2024 via email. The notice explained the reason for the recall and requested the following: "ACTIONS REQUIRED: 1. Examine your inventory and quarantine the affected product codes and lots subject to this recall action. 2. Distributors, please identify your customers, and notify them at once of this recall action to the distributor level. Such notification may be enhanced by including the enclosed letter (Appendix B) for your customers and Healthcare Providers (HCP s) in the field. 3. Return the enclosed Appendix B with the information via email to recalls@sol-m.com. The quarantine should be carried out to the wholesale and distributor level. Products in the field beyond the control of the distributor or wholesaler should be recalled from the field and returned to the distributor. Distributors should quarantine the products in inventory and quarantine the products received from customers until the 510(k) submissions are pending or cleared. The returned product should be quarantined in a secure location to avoid distribution based on your quality management procedures. We will provide communication with any updates as we work with the FDA."
Quantity
not on file
Distribution
US Nationwide, Czech Republic, Belgium, Portugal, Canada, Saudi Arabia, Netherland, France, Switzerland, Latvia, UK, Spain, Germany, Philippines, Australia, Malaysia, Jamaica, Slovenia, Brazil, Taiwan, Denmark, Sweden, Singapore, Romania, Poland, Barbados, Mexico, Guatemala, Chile, Ireland, Italy, Panama
Lot, serial or model codes
a) Model No. 180011LDS, UPC 840368511899, Lot Numbers: 1102070, 1107062, 2002005, 2006004, 2012008, 2909022, 2911053, 19052401; b) Model No. P180011LDS, UPC 810018097554, Lot Numbers: 4104025, 4104100, 4108014, 4108015, 4108017, 4108095, 4109088, 4202001, 4301003, 4303047, 4312032; c) Model No. 181023L, UPC Sold in Canada Only, Lot Numbers: 4301029; d) Model No. 12558LDS, UPC 810062435319, Lot Numbers: 7210016; e) Model No. P180001PP, UPC 818392018165, Lot Numbers: 4104027, 4104028, 4104029, 4104030, 4104031, 4104032, 4105038, 4106031, 4106032, 4106033, 4106074, 4106075, 4106079, 4106080, 4106081, 4106082, 4106083, 4106086, 8103004, 8104004, 8104006, 8104007, 8104008, 8105038, 8105047, 8105053, 8105054, 8106011, 8106014, 8106024, 8106025, 8106026, 8106027, 8106028, 8106029, 8106030, 8106031, 8106032, 8106033; f) Model No. 181025S, UPC NONE, Lot Numbers: 4110059, 4110091, 4110092, 4212021, 4303009, 4303010, 4303011, 4303012, 4304016,
510(k) numbers
not on file
PMA numbers
not on file
Product code
FMF
Firm FEI
3014312726
Firm city
Chicago
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28