Medline Industries, LP: Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators C-SECTION PREP DYNJ32702B
- Recall number
- Z-3077-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators C-SECTION PREP DYNJ32702B
- Status
- Open, Classified
- Initiated
- 2026-07-20
- Posted
- 2026-08-31
- Terminated
- not on file
- Reason
- Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
- Root cause
- Nonconforming Material/Component
- Action
- Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
- Quantity
- 222,653 total
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- UDI-DI 10193489431070 Lots 25BBB005 25GBY063 25KBI862 25LBT688 26CBG211
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OHM
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28