Skip to the content

FDA Medical Device Recalls (openFDA)

Edwards Lifesciences, LLC: Fogarty Adherent Clot Catheter Model Numbers 1408010 and 140806 Product Usage: The Fogarty adherent clot catheters are designed to effectively remove clot material in the peripheral vasculature. The catheters are indicated for the removal of emboli and thrombi from native arteries or synthetic grafts. The device are designed to facilitate clots too resistant to be removed by an elastomeric balloon. The device features a spiral-shaped, latex-covered stainless steel cable that assumes a corkscrew shape when retracted, greatly expanding the surface areas to entrap fibrous material. The device is 80 cm in length and is available in 4F-6F diameters.

Recall number
Z-3077-2017
Recalling firm
Edwards Lifesciences, LLC
Product
Fogarty Adherent Clot Catheter Model Numbers 1408010 and 140806 Product Usage: The Fogarty adherent clot catheters are designed to effectively remove clot material in the peripheral vasculature. The catheters are indicated for the removal of emboli and thrombi from native arteries or synthetic grafts. The device are designed to facilitate clots too resistant to be removed by an elastomeric balloon. The device features a spiral-shaped, latex-covered stainless steel cable that assumes a corkscrew shape when retracted, greatly expanding the surface areas to entrap fibrous material. The device is 80 cm in length and is available in 4F-6F diameters.
Status
Terminated
Initiated
2017-07-06
Posted
not on file
Terminated
2018-10-17
Reason
The diameter of the Fogarty Catheter balloon, measures 6mm instead of 10mm.
Root cause
Packaging
Action
Recall notification letters were sent to affected customers on 7/6/2017. The recall notice included a response form and instructed consignee to return any unused devices..
Quantity
200 units
Distribution
Worldwide Distribution - US Nationwide GB, NL, DE, IT, FR, BE, ES, NO, CZ, BG, SE, AT, CH, PL, RU, FI
Lot, serial or model codes
Lots 60439817 and 60439818
510(k) numbers
["K901625"]
PMA numbers
not on file
Product code
DXE
Firm FEI
2015691
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register