Skip to the content

FDA Medical Device Recalls (openFDA)

Biomerieux Inc: VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.

Recall number
Z-3075-2017
Recalling firm
Biomerieux Inc
Product
VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.
Status
Terminated
Initiated
2017-07-12
Posted
not on file
Terminated
2020-12-16
Reason
Customers have reported that some VITEK 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.
Root cause
Other
Action
Consignees (Subsidiaries) of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on 07/12/2017. The direct consignees (subsidiaries and distributors) have until 08/12/2017 to notify their respective impacted users. The directions to the user are as follows: Please implement the following actions at this time: " Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. " Refrain from performing the VITEK 2 Systems Software 8.01 update. o A solution has been identified, and your local bioMrieux representative will contact you with further instruction in the coming weeks. " If you have updated VITEK 2 Systems Software to version 8.01 within the last seven (7) days, please contact your local bioMrieux representative. " Please store this letter with your bioMrieux instrument documentation. " Complete the Acknowledgement Form and return it to your local bioMrieux representative."
Quantity
1866 units
Distribution
Worldwide
Lot, serial or model codes
VITEK¿ 2 Compact 15 and Compact 30 instruments previously manufactured/installed with VITEK¿ 2 System software versions prior to 5.01, and currently being upgraded to version 8.01.
510(k) numbers
["K050002"]
PMA numbers
not on file
Product code
LON
Firm FEI
1950204
Firm city
Hazelwood
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register