FDA Medical Device Recalls (openFDA)
Medspira, Llc: mcompass 2-channel Balloon Catheters The mcompass Biofeedback Anorectal Manometry System is for use on patients requiring anorectal pressure studies. mcompass Anorectal Manometry System (RMD-001-001) which includes a FOB component connected to the balloon catheter to inflate and deflate and a medical grade tablet PC to run the software. The software is not included in the recall, only the balloon catheter component.
- Recall number
- Z-3072-2017
- Recalling firm
- Medspira, Llc
- Product
- mcompass 2-channel Balloon Catheters The mcompass Biofeedback Anorectal Manometry System is for use on patients requiring anorectal pressure studies. mcompass Anorectal Manometry System (RMD-001-001) which includes a FOB component connected to the balloon catheter to inflate and deflate and a medical grade tablet PC to run the software. The software is not included in the recall, only the balloon catheter component.
- Status
- Terminated
- Initiated
- 2017-06-14
- Posted
- not on file
- Terminated
- 2021-08-24
- Reason
- Potential failure in the balloon bond in the inner catheter stem, leading to the separation of the balloon from the inner stem, leaving it in the rectal cavity.
- Root cause
- Employee error
- Action
- The firm Medspira plans to contact all foreign and domestic consignees in regards to the recall. International consignees are to destroy recalled product by cutting the catheter shaft and balloon with scissors. It is also requested that their own customers be notified of the recall. Medspira plans to provide product replacement. Domestic customers are asked to return affected catheters to the recalling firm. Medspira indicates that they will cove shipping costs and replacements. All consignees are asked to complete and return the attached recall response form. For further questions please call, (612) 789-0013.
- Quantity
- 355 catheters (255 US - 100 foreign. )
- Distribution
- Worldwide Distribution - US Distribution to the states of :CA, FL, IA, ID, and TX., and to the countries of : Austria, Italy, Malaysia, Philippines, Sweden, and United Kingdom
- Lot, serial or model codes
- Part # RMD-002-004 Lot No #'s 160627-10 and 160627-11.
- 510(k) numbers
- ["K120088","K143031"]
- PMA numbers
- not on file
- Product code
- GBT
- Firm FEI
- 3008414621
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28