FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators IV START KIT LAB DRAW-SH CDS860014Q IV START KIT DYKM1000B IV START KIT LAB DRAW-SH CDS860014P ULTRASOUND GUIDED PIV START KI DYNDV2684
- Recall number
- Z-3071-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators IV START KIT LAB DRAW-SH CDS860014Q IV START KIT DYKM1000B IV START KIT LAB DRAW-SH CDS860014P ULTRASOUND GUIDED PIV START KI DYNDV2684
- Status
- Open, Classified
- Initiated
- 2026-07-20
- Posted
- 2026-08-31
- Terminated
- not on file
- Reason
- Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
- Root cause
- Nonconforming Material/Component
- Action
- Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
- Quantity
- 222,653 total
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- CDS860014Q UDI-DI 10198459629822 Lots 26ABR359 26BBP789 26CBL940 26DBK987 26FBK255 26FBK816 DYKM1000B UDI-DI 10889942508143 Lots 24KMD782 24LMH404 25AMD765 25BMD744 CDS860014P UDI-DI 10888277797819 Lots 24IBA155 24JBY045 24LBM014 25HBR509 25JBX951 DYNDV2684 UDI-DI 10198459185465 Lot 24LBP243
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRS
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28