FDA Medical Device Recalls (openFDA)
Toshiba American Medical Systems Inc: Toshiba America Medical Systems Ultimax Digital Fluoroscopic X-Ray System Model: DREX-UI80 Ultimax is intended for gastrointestinal studies, vascular studies, general radiography, and fluoroscopy. The system has medical applications ranging from but not limited to: contrast-enhanced studies, support of endoscopic studies, nonvascular contrast-enhanced studies, nonvascular IVR, vascular contrast-enhanced studies support of vascular NR, and general radiography.
- Recall number
- Z-3070-2017
- Recalling firm
- Toshiba American Medical Systems Inc
- Product
- Toshiba America Medical Systems Ultimax Digital Fluoroscopic X-Ray System Model: DREX-UI80 Ultimax is intended for gastrointestinal studies, vascular studies, general radiography, and fluoroscopy. The system has medical applications ranging from but not limited to: contrast-enhanced studies, support of endoscopic studies, nonvascular contrast-enhanced studies, nonvascular IVR, vascular contrast-enhanced studies support of vascular NR, and general radiography.
- Status
- Terminated
- Initiated
- 2017-03-15
- Posted
- not on file
- Terminated
- 2020-09-25
- Reason
- The issue was identified as a firmware problem with the generator.
- Root cause
- Software design
- Action
- Toshiba America Medical Systems ( TAMS) will bring the defect into Compliance: 1.You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2.You will install modified firmware on the systems to correct the defect and prevent recurrence. 3.You will provide progress reports on the effectiveness of the plan. For further questions, please call (714) 730-5000.
- Quantity
- 18
- Distribution
- US Distribution to the states of : LA, AR, IA, TX, IL, MN. MS, NJ, WV, SC, LA, WY, NC and CA.
- Lot, serial or model codes
- Ser. No. U1A1692606
- 510(k) numbers
- ["K170832"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z