Skip to the content

FDA Medical Device Recalls (openFDA)

Arjo, Inc. dba ArjoHuntleigh: Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a swimming/therapy pool; Arjo Hospital Equipment AB, Eslov, Sweden; Models GAB1000-01-xx, GAB1000-11-xx and GAC100x Entroy is a pool lift that docks to either a chair or a stretcher and aids disabled people getting in and out of a pool. This equipment is intended for disabled persons who need to transfer in and out of a therapy and swimming pool under supervision of skilled and trained personnel in accordance with the Entroy Operating and Product Care Instructions. All other use must be avoided. Active/semi-active persons can and should use the Entroy chair (with or without chassis) while bedridden persons without trunk stability can and should be transferred with the Entroy stretcher with chassis.

Recall number
Z-3070-2011
Recalling firm
Arjo, Inc. dba ArjoHuntleigh
Product
Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a swimming/therapy pool; Arjo Hospital Equipment AB, Eslov, Sweden; Models GAB1000-01-xx, GAB1000-11-xx and GAC100x Entroy is a pool lift that docks to either a chair or a stretcher and aids disabled people getting in and out of a pool. This equipment is intended for disabled persons who need to transfer in and out of a therapy and swimming pool under supervision of skilled and trained personnel in accordance with the Entroy Operating and Product Care Instructions. All other use must be avoided. Active/semi-active persons can and should use the Entroy chair (with or without chassis) while bedridden persons without trunk stability can and should be transferred with the Entroy stretcher with chassis.
Status
Terminated
Initiated
2011-08-04
Posted
2011-08-24
Terminated
2011-11-28
Reason
The Entroy Pool Lift boom may rapidly lower un-commanded to the bottom of its stroke, with a potential for injury to the resident or caregiver.
Root cause
Device Design
Action
ArjoHuntleigh sent an Urgent Field Correction Notice I.D. 110701AHE to the affected end user via UPS 2nd Day on August 4, 2011. The notice informed the account of the potential for lift boom to rapidly lower un-commanded to the bottom of its stroke. The account was requested to immediately cease use of the affected lift, and disconnect the power to the lift to prevent further use until corrected. The account was requested to complete and sign the Customer Response Form, and return it to ArjoHuntleigh to schedule a service call to correct the lift. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 54485.
Quantity
1 unit in the U.S., 311 units in the rest of the world
Distribution
Worldwide Distribution - USA including CT
Lot, serial or model codes
Model GAB1000-01-US, serial number SEE0326028  Rest of the world: Models GAB1000-01-xx, GAB1000-11-xx and GAC100x, all serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
1419652
Firm city
Addison
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register