Skip to the content

FDA Medical Device Recalls (openFDA)

Ultradent Products, Inc.: Brand Name: Transcend Product Name: Transcend Singles Universal Body (UB) Refill 10pk Model/Catalog Number: 4757- Software Version: N/A Product Description: Transcend universal composite is a light-cured, tooth shade resin composite material (Bis-GMA-based) to be used for direct and indirect posterior and anterior tooth restorations. Transcend universal composite is radiopaque and available in a range of dentin, enamel, and body shades. It is 79% filled by weight and 60-61% filled by volume and has an average particle size of 0.9 m (by weight) with narrow upper limit particle distribution. Component: N/A

Recall number
Z-3068-2024
Recalling firm
Ultradent Products, Inc.
Product
Brand Name: Transcend Product Name: Transcend Singles Universal Body (UB) Refill 10pk Model/Catalog Number: 4757- Software Version: N/A Product Description: Transcend universal composite is a light-cured, tooth shade resin composite material (Bis-GMA-based) to be used for direct and indirect posterior and anterior tooth restorations. Transcend universal composite is radiopaque and available in a range of dentin, enamel, and body shades. It is 79% filled by weight and 60-61% filled by volume and has an average particle size of 0.9 m (by weight) with narrow upper limit particle distribution. Component: N/A
Status
Open, Classified
Initiated
2024-08-02
Posted
2024-09-06
Terminated
not on file
Reason
Due to a manufacturing issue, red particulates present in the expressed composite.
Root cause
Material/Component Contamination
Action
On 08/6/2024, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via 3-day delivery to customer informing them that firm has identified a potential presence of red or blue polyvinyl particulate in the composite. Customer are instructed to: -Discard all remaining Transcend Singles or Vit-lescence Singles from lots listed in the Recall Letter. -Fill out Customer Response Form and return to Ultradent indicating disposal of product. -For any concerns or questions related to your replacement product or product use, contact Ultradent Customer Service at 1-888-230-1420. -Should you experience any adverse reactions or quality problems with the use of this product, these can also be reported to the FDA s MedWatch Adverse Event Reporting program with online, by regular mail or by fax. o Complete and submit the report online o Regular mail or fax: download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Quantity
2450
Distribution
US: AZ, CA,CO,FL,GA,HI, IA, ID,IL,IN,KS,KY,LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, SC, TN, TX, UT, VA, VT, WI, WV Foreign: NL, DE, CH, CZ, FR, ES
Lot, serial or model codes
Lot Code: Model No 4757; UDI-DI 00883205028334; Lot Number C15C2; Expiration Date 31-May-2026
510(k) numbers
["K201795"]
PMA numbers
not on file
Product code
EBF
Firm FEI
1718912
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register