Ultradent Products, Inc.: Brand Name: Transcend Product Name: Transcend Singles Universal Body (UB) Refill 10pk Model/Catalog Number: 4757- Software Version: N/A Product Description: Transcend universal composite is a light-cured, tooth shade resin composite material (Bis-GMA-based) to be used for direct and indirect posterior and anterior tooth restorations. Transcend universal composite is radiopaque and available in a range of dentin, enamel, and body shades. It is 79% filled by weight and 60-61% filled by volume and has an average particle size of 0.9 m (by weight) with narrow upper limit particle distribution. Component: N/A
- Recall number
- Z-3068-2024
- Recalling firm
- Ultradent Products, Inc.
- Product
- Brand Name: Transcend Product Name: Transcend Singles Universal Body (UB) Refill 10pk Model/Catalog Number: 4757- Software Version: N/A Product Description: Transcend universal composite is a light-cured, tooth shade resin composite material (Bis-GMA-based) to be used for direct and indirect posterior and anterior tooth restorations. Transcend universal composite is radiopaque and available in a range of dentin, enamel, and body shades. It is 79% filled by weight and 60-61% filled by volume and has an average particle size of 0.9 m (by weight) with narrow upper limit particle distribution. Component: N/A
- Status
- Open, Classified
- Initiated
- 2024-08-02
- Posted
- 2024-09-06
- Terminated
- not on file
- Reason
- Due to a manufacturing issue, red particulates present in the expressed composite.
- Root cause
- Material/Component Contamination
- Action
- On 08/6/2024, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via 3-day delivery to customer informing them that firm has identified a potential presence of red or blue polyvinyl particulate in the composite. Customer are instructed to: -Discard all remaining Transcend Singles or Vit-lescence Singles from lots listed in the Recall Letter. -Fill out Customer Response Form and return to Ultradent indicating disposal of product. -For any concerns or questions related to your replacement product or product use, contact Ultradent Customer Service at 1-888-230-1420. -Should you experience any adverse reactions or quality problems with the use of this product, these can also be reported to the FDA s MedWatch Adverse Event Reporting program with online, by regular mail or by fax. o Complete and submit the report online o Regular mail or fax: download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
- Quantity
- 2450
- Distribution
- US: AZ, CA,CO,FL,GA,HI, IA, ID,IL,IN,KS,KY,LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, SC, TN, TX, UT, VA, VT, WI, WV Foreign: NL, DE, CH, CZ, FR, ES
- Lot, serial or model codes
- Lot Code: Model No 4757; UDI-DI 00883205028334; Lot Number C15C2; Expiration Date 31-May-2026
- 510(k) numbers
- ["K201795"]
- PMA numbers
- not on file
- Product code
- EBF
- Firm FEI
- 1718912
- Firm city
- South Jordan
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28