Pentax of America Inc: Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.

Recall number
Z-3067-2017
Recalling firm
Pentax of America Inc
Product
Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Status
Terminated
Initiated
2017-05-23
Posted
not on file
Terminated
2020-05-21
Reason
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
Root cause
Error in labeling
Action
Recall Notifications were sent to 104 consignees via USPS Certified Mail on 5/23/2017. Included with the Field Safety letter was the customer response form, and the addendum to IFU Z845. These were also posted on Pentax of America web page until the recall is closed.
Quantity
US: 118 units
Distribution
Nationwide
Lot, serial or model codes
Serial Numbers: G110307, H110482, H110484, H110487, H110489, G110334, G110338, G110404, G110296, G110358, G110340, G110341, G110407, G110409, H110432, G110345, H110450, G110391, H110476, H110443, H110446, G110400, H110483, G110394, G110406, H110411,  H110497, G110360, G110363, G110364, G110371, G110366, G110368,  G110372, H110434, H110447, G110380, H110455, G110298, H110445,  H110449, H110461, G110292, G110346, G110365, H110494, H110495,  H110474, G110376, G110319, G110328, H110473, H110413, H110414, H110427, H110453, H110454, H110415, G110312, G110321, H110467,  H110471, H110412, G110289, G110303, G110294, H110420, G110299,  G110300, H110448, H110452, G110386, G110331, H110451, H110493,  G110313, G110314, H110464, H110496, H110433, G110373, G110375,  G110317, H110431, H110417, G110291, G110302, G110369, G110378,  G110335, G110387, G110390, G110392, G110395, G110311, G110327,  G110332, G110382, G110355, G110357, G110287, G110320, H110435,  H110481, H110485, H110486, G110354, G110316, G110318, G110362,  G110393, H110498, H110456, H110457, H110459, G110337, G110356,  H110421, H110429, G110290, G110352, H110470, H110478, G110383,  G110384, G110361.
510(k) numbers
["K061068","K090196","K130706"]
PMA numbers
not on file
Product code
ODG
Firm FEI
1000080301
Firm city
Montvale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register