Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11319151

Recall number
Z-3064-2024
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11319151
Status
Open, Classified
Initiated
2024-07-29
Posted
2024-09-06
Terminated
not on file
Reason
The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.
Root cause
Under Investigation by firm
Action
On July 29, 2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Impact to Results Erroneously depressed or elevated urine protein or cerebrospinal fluid patient results may occur. Siemens investigation revealed a positive bias of up to 52% (at 27.3 mg/dL (273 mg/L); up to 19% at 68.7 mg/dL (687 mg/L)). However, analyte recovery decreases rapidly after opening a reagent pack and a negative bias of up to -35% was observed at 24 hours (at 27 mg/dL (270 mg/L); up to -9% at 67 mg/dL (670 mg/L)). Results of this assay would be interpreted in conjunction with the patient s medical history, clinical presentation, and other findings. Customer Actions: 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Discontinue use of and discard the kit lot listed in the table above (Products Section). 3. Complete and return the Field Correction Effectiveness Check and indicate product replacement needs on the form attached to this letter within 30 days. 4. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Quantity
143 units (17 US, 126 OUS)
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, IL, KY, LA, MS, NJ, NY, NV, PA, TX, VA, PR and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Columbia, Croatia, Cura¿ao, St Eus, Czech Republic, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, India, Ireland, Italy, Jordan, Kenya, Korea, Kuwait, Latvia, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E., United Kingdom, Uruguay, Vatikancity.
Lot, serial or model codes
Siemens Material Number: 11319151; UDI-DI: 00630414279176; Lot number: 140
510(k) numbers
not on file
PMA numbers
not on file
Product code
CEK
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register