Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.

Recall number
Z-3062-2024
Recalling firm
Bard Peripheral Vascular Inc
Product
Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.
Status
Open, Classified
Initiated
2024-07-10
Posted
2024-09-06
Terminated
not on file
Reason
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.
Root cause
Under Investigation by firm
Action
On 08/01/2024, the firm sent an "URGENT: Medical Device Product Removal" letter via US mail and email to customers informing them that the firm has received multiple reports that guidewire may have been mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Customers are instructed to: 1. Discontinue use of product listed in the letter, and immediately dispose of all affected product remaining in their possession in accordance with their local facility s process. 2. Share this notification with all users within their facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, ensure that these entities are contacted and informed of the recall. 3. Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not they have any of the product so that BD may acknowledge your receipt of this notification and subsequently process any required credits. 4. Indicate on the response form the quantity of product identified at their facility and confirm that this product inventory was disposed of. 5. If products were purchased from a distributor, contact the distributor for further instructions and credit resolution. For questions/assistance contact: BD North American Regional Complaint Center for Product Complaints/Technical Questions at: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Complaints when prompted Mon Fri 8:00am and 5:00pm CT, or Email: productcomplaints@bd.com. For recall related questions email BDRC14@BD.COM
Quantity
625 units
Distribution
Worldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The countries of Belgium, Canada, France, Italy, and United Kingdom.
Lot, serial or model codes
Catalog Number: 5583705 Lot Number / UDI codes: REGZ2564 (01)00801741085451(17)271130(10)REGZ2564 REFV0870 (01)00801741085451(17)260731(10)REFV0870 REFU4050 (01)00801741085451(17)260630(10)REFU4050 REFT3309 (01)00801741085451(17)260531(10)REFT3309 REEX1002 (01)00801741085451(17)250930(10)REEX1002 REEW2740 (01)00801741085451(17)250831(10)REEW2740 REEV0897 (01)00801741085451(17)250731(10)REEV0897 REDY2576 (01)00801741085451(17)241031(10)REDY2576 REEU1584 (01)00801741085451(17)250630(10)REEU1584 REER0990 (01)00801741085451(17)250331(10)REER0990 REEQ0825 (01)00801741085451(17)250228(10)REEQ0825
510(k) numbers
["K770977"]
PMA numbers
not on file
Product code
DQX
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register