FDA Medical Device Recalls (openFDA)
Maquet Inc.: Maquet HL Console for Heart Lung Machine HL20: HL 20 4-pumps console base: P/N 917741 HL 20 5-pumps console base: P/N 917751 The HL 20 is indicated for use as an extracorporeal circulation device for perfusion lasting not more then six hours. The HL console is equipped with a rack for accommodating the system's electronic modules.
- Recall number
- Z-3057-2011
- Recalling firm
- Maquet Inc.
- Product
- Maquet HL Console for Heart Lung Machine HL20: HL 20 4-pumps console base: P/N 917741 HL 20 5-pumps console base: P/N 917751 The HL 20 is indicated for use as an extracorporeal circulation device for perfusion lasting not more then six hours. The HL console is equipped with a rack for accommodating the system's electronic modules.
- Status
- Terminated
- Initiated
- 2011-06-20
- Posted
- 2011-08-22
- Terminated
- 2012-10-16
- Reason
- HL 20 Consoles for 4-pump and 5-pump configuration can malfunction if an excessive amount of liquid is used causing circuit board malfunction.
- Root cause
- Device Design
- Action
- Maquet sent an URGENT PRODUCT RECALL NOTIFICATION letter dated June 17, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to acknowledge receipt of the notification by completing and sending back the attached response form whether or not they have any affected product, fax to 973-807-9210 no later than June 30, 2011. For any questions customers were instructed to contact thier local MAQUET representative.
- Quantity
- 229 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- 510K numer K943803 Device Listing #E614811 HL-20 Console
- 510(k) numbers
- ["K943803"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 3008355164
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28