FDA Medical Device Recalls (openFDA)
AVID Medical, Inc.: Halyard medical convenience kits labeled as: 1) A-P PACK - 205936, Model/Item Number (kit) ANAP22-01; 2) POSTERIOR PACK - 205960, Model/Item Number (kit) ANPP15-01; 3) ANEURYSM PACK, Model/Item Number (kit) BPAY14-01; 4) SPINE PACK, Model/Item Number (kit) HSLM53-01; 5) ANTERIOR CERVICAL PACK - 205988, Model/Item Number (kit) MHAC65-01; 6) SPINE PACK - 206013, Model/Item Number (kit) MHSP99-01; 7) CRANI PACK, Model/Item Number (kit) WECR41-01.
- Recall number
- Z-3046-2026
- Recalling firm
- AVID Medical, Inc.
- Product
- Halyard medical convenience kits labeled as: 1) A-P PACK - 205936, Model/Item Number (kit) ANAP22-01; 2) POSTERIOR PACK - 205960, Model/Item Number (kit) ANPP15-01; 3) ANEURYSM PACK, Model/Item Number (kit) BPAY14-01; 4) SPINE PACK, Model/Item Number (kit) HSLM53-01; 5) ANTERIOR CERVICAL PACK - 205988, Model/Item Number (kit) MHAC65-01; 6) SPINE PACK - 206013, Model/Item Number (kit) MHSP99-01; 7) CRANI PACK, Model/Item Number (kit) WECR41-01.
- Status
- Open, Classified
- Initiated
- 2026-05-19
- Posted
- 2026-08-28
- Terminated
- not on file
- Reason
- The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
- Root cause
- Material/Component Contamination
- Action
- Owens & Minor issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION to its consignees on 5/19/2026 via email. The notice explained the issue, potential risk to health and requested the following: "ACTIONS TO TAKE: " Identify, segregate, and quarantine all impacted product. " Add recall labels to affected Avid kits to ensure all users are aware of the actions to discard the impacted component. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness."
- Quantity
- 816 kits
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, WA, WY, TX, MI, FL, MD, UT, NC, MS, PA, MO, IA, NE, OH, IL, SD, MN, GA.
- Lot, serial or model codes
- Model/Item Number (kit) ANAP22-01, UDI/DI 10809160473146, Kit Lot Number 346967; Model/Item Number (kit) ANAP22-01, UDI/DI 10809160473146, Kit Lot Number 346982; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343413; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343414; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343713; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343715; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343716; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343412; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343416; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343714; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343717; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343718; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343719; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343723; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 343746; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 343747; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 343771; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 374128; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 374129; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 374296; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 342005; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 344071; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 347906; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 341835; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 343206; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 343208; Model/Item Number (kit) MHAC65-01, UDI/DI 10809160479537, Kit Lot Number 345289; Model/Item Number (kit) MHSP99-01, UDI/DI 10809160479797, Kit Lot Number 345694; Model/Item Number (kit) MHSP99-01, UDI/DI 10809160479797, Kit Lot Number 341654; Model/Item Number (kit) WECR41-01, UDI/DI 10809160485392, Kit Lot Number 344464.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJH
- Firm FEI
- 1047429
- Firm city
- Toano
- Firm state
- VA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28