Ameditech Inc: ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1000/OPI2000)+AMP+BZO+COC+THC", PSD-6MTDBO-300 /"ProScreen" Drugs of Abuse Test (COC/THC)+BZO+MTD+OPI300+OXY", PSCUP-5M-W/ ProScreen" Drugs of Abuse Cup AMP/COC/MET1000/OPI2000/THC), PSD-10MOB "ProScreen" Drugs of Abuse Test (COC/THC)+(MET1000/OPI2000)+(OXY/ MDMA)+AMP+BZO+BUP+MTD", PSD-5M / "ProScreen" Drugs of Abuse Test AMP/COC/MET1000/OPI2000/THC", PSCUPA-10OM / ProScreen" Drugs of Abuse Cup (COC/THC)+(MET1000/OPI2000)+(OXY/ MDMA) +(PCP/AMP)+BAR+BZO
- Recall number
- Z-3042-2020
- Recalling firm
- Ameditech Inc
- Product
- ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1000/OPI2000)+AMP+BZO+COC+THC", PSD-6MTDBO-300 /"ProScreen" Drugs of Abuse Test (COC/THC)+BZO+MTD+OPI300+OXY", PSCUP-5M-W/ ProScreen" Drugs of Abuse Cup AMP/COC/MET1000/OPI2000/THC), PSD-10MOB "ProScreen" Drugs of Abuse Test (COC/THC)+(MET1000/OPI2000)+(OXY/ MDMA)+AMP+BZO+BUP+MTD", PSD-5M / "ProScreen" Drugs of Abuse Test AMP/COC/MET1000/OPI2000/THC", PSCUPA-10OM / ProScreen" Drugs of Abuse Cup (COC/THC)+(MET1000/OPI2000)+(OXY/ MDMA) +(PCP/AMP)+BAR+BZO
- Status
- Terminated
- Initiated
- 2020-08-18
- Posted
- not on file
- Terminated
- 2024-09-30
- Reason
- Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
- Root cause
- Process control
- Action
- A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
- Quantity
- 51,075 units
- Distribution
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
- Lot, serial or model codes
- ALL Lots Within Expiry
- 510(k) numbers
- ["K052115","K113046"]
- PMA numbers
- not on file
- Product code
- DKC
- Firm FEI
- 3003789989
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28