Skip to the content

FDA Medical Device Recalls (openFDA)

Ameditech Inc: MicroTox (AMT Item Number/ AMT Description): MTPA-6MBAU / "MicroTox (AMP300/BZO200)+(COC300/THC50)+ (MET300/MOR300) With Cr, Ni, pH, Bl, S.G.", MTPA-6MBAU/ "MicroTox (AMP300/BZO200)+(COC300/THC50)+ (MET300/MOR300) With Cr, Ni, pH, Bl, S.G."

Recall number
Z-3038-2020
Recalling firm
Ameditech Inc
Product
MicroTox (AMT Item Number/ AMT Description): MTPA-6MBAU / "MicroTox (AMP300/BZO200)+(COC300/THC50)+ (MET300/MOR300) With Cr, Ni, pH, Bl, S.G.", MTPA-6MBAU/ "MicroTox (AMP300/BZO200)+(COC300/THC50)+ (MET300/MOR300) With Cr, Ni, pH, Bl, S.G."
Status
Terminated
Initiated
2020-08-18
Posted
not on file
Terminated
2024-09-30
Reason
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Root cause
Process control
Action
A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
Quantity
12, 100 units
Distribution
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Lot, serial or model codes
ALL Lots Within Expiry
510(k) numbers
not on file
PMA numbers
not on file
Product code
DKC
Firm FEI
3003789989
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register