Skip to the content

FDA Medical Device Recalls (openFDA)

Magellan Diagnostics, Inc.: LeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Product Usage: IVD test kit for the detection of lead in blood. The kit consists of sensors, treatment reagent, lead controls (two levels), heparinized capillary tubes and plungers, transfer droppers and a calibration button. The LeadCare II Lead Controls packaged in the kit lots listed in #2 below are the subject of this recall (Field Correction) action.

Recall number
Z-3035-2017
Recalling firm
Magellan Diagnostics, Inc.
Product
LeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Product Usage: IVD test kit for the detection of lead in blood. The kit consists of sensors, treatment reagent, lead controls (two levels), heparinized capillary tubes and plungers, transfer droppers and a calibration button. The LeadCare II Lead Controls packaged in the kit lots listed in #2 below are the subject of this recall (Field Correction) action.
Status
Terminated
Initiated
2016-01-20
Posted
2017-08-15
Terminated
2018-02-21
Reason
LeadCare II Test Kits controls out of range
Root cause
Device Design
Action
Magellan notified customers initially by Email, fax and/or verbal communication with Magellan Product Support representatives beginning on January 20, 2016 to advise of the problem and provide revised Assignment Control Value Matrix..
Quantity
12,931 kits
Distribution
Worldwide - US Nationwide distribution and the countries of: COLOMBIA, CANADA, THAILAND, INDIA, UAE, AUSTRALIA, GERMANY, CANADA, SOUTH AFRICA, SPAIN, NEW ZEALAND, KOREA
Lot, serial or model codes
Lot # :1507N Exp Date: 1/27/17; Lot # :1508N Exp Date: 2/27/17; Lot # :1510N Exp Date: 4/19/17; Lot # :1511M Exp Date: 5/3/17 Lot # :1511N Exp Date: 5/17/17 Lot # :1512M Exp Date: 6/2/17
510(k) numbers
["K052549"]
PMA numbers
not on file
Product code
DOF
Firm FEI
1218996
Firm city
North Billerica
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register