Skip to the content

FDA Medical Device Recalls (openFDA)

FUJIFILM Healthcare Americas Corporation: FUJIFILM Synapse PACS software v7.4.310; radiological image processing system

Recall number
Z-3030-2026
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
FUJIFILM Synapse PACS software v7.4.310; radiological image processing system
Status
Open, Classified
Initiated
2026-07-22
Posted
2026-08-26
Terminated
not on file
Reason
Software issue may cause the ruler measurement to be inaccurate.
Root cause
Software design
Action
FUJIFILM issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 7/21/2026 via email. The notice explained the issue and how to identify it, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1. Please post this notification on or near the Synapse PACs workstation. 2. Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 3. Please contact Technical Assistance Center (TAC) at 1-888-FUJIMED (1-888-385-4633) to open a service ticket. 4. Please report any known adverse events to FDA via MedWatch. Regarding this issue, Synapse PACS can safely be used by the end-user as long as: 5. The end-user is aware of the issue and, 6. Appropriate actions are taken when rendering a diagnosis for patients. The following workarounds can be used: 7. Perform M-mode ruler measurements on the ultrasound system and then transfer the study to Synapse PACS." For questions regarding this notice, please email hcusquality@fujifilm.com.
Quantity
6 units
Distribution
Worldwide - US Nationwide distribution in the states of NY, VA, KS, OR, OH, and the country of Australia.
Lot, serial or model codes
UDI (01)00854904006008(10)0704310, Serial Numbers: CV-00272120, CV-0080080176, CV-01318490, CV-01-80102732, CV-80057523, CV-00494020-38
510(k) numbers
["K190232"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000513161
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register