GE Medical Systems, LLC: Optima MR750 3.0T Product Usage: The Optima MR450 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Optima MR450 1.5T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Recall number
Z-3023-2017
Recalling firm
GE Medical Systems, LLC
Product
Optima MR750 3.0T Product Usage: The Optima MR450 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Optima MR450 1.5T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Status
Terminated
Initiated
2017-05-11
Posted
2017-08-02
Terminated
2020-09-09
Reason
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Root cause
Mixed-up of materials/components
Action
Consignees received a written recall notice dated 05/17/2017 notifying them of a correction to all units using an Uninterruptable Power Supply. Consignees were instructed not to unplug the unit and to continue normal use. Units with the Eaton 9130 Uninterruptable Power Supply connected to the MR operator console would receive a correction at no charge. This would involve possibly replacing the power cable. GE Representatives will be contacting consignees to arrange corrections. Any questions can be directed to GE Healthcare Service at 1-800-437-1171 or your local service representative.
Quantity
68 units in total
Distribution
US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
Lot, serial or model codes
Medical Device Listing No. D117354, D295466, D182389, D102432, D206140, D230310, D269569, D102430, D140572, D188736, D102434, D234841, D196247, D134019, D236543, D006275, D159181, D288890, D229859; Model No. Eaton 9130, 5323286, 5396502-5, 5499460-3, 5499460-6, 5499460-9, 5741000-3, 00840682115971, 5396502-2, 5499460-2, 5741000-2, 00840682115872, 5373011-3, 5396469-3, 5481490-2, 5481500-7, 5499460-7, 5499460-10, 5741000, 00840682103817, 5323286-101, 2377062-61, 2395001, 2395001-2, 2395001-3, 2395001-4, 00840682105699, 840682105378
510(k) numbers
["K083147","K132376","K142085","K160618"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2183553
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register