Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Olympus Medical Systems Corporation (OMSC) Evis Exera II Bronchovideoscope Olympus BF Type Q180 (BF-Q180) All serial numbers - Product Usage: designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the airways and tracheobronchial tree.

Recall number
Z-3017-2020
Recalling firm
Olympus Corporation of the Americas
Product
Olympus Medical Systems Corporation (OMSC) Evis Exera II Bronchovideoscope Olympus BF Type Q180 (BF-Q180) All serial numbers - Product Usage: designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the airways and tracheobronchial tree.
Status
Terminated
Initiated
2020-08-31
Posted
not on file
Terminated
2024-02-14
Reason
The BF-Q180 bronchoscope is being recalled because it does not have a 510(k) clearance, and the device is associated with a higher rate of patient infections than other comparable OMSC bronchoscopes.
Root cause
Device Design
Action
Olympus Corporation notified direct customer accounts on 08/31/2020 via recall notification letter. Letter states reason for recall, health risk, and action to take. Actions to be taken by customers include the following: 1. Inspect your inventory and identify any BF-Q180 models. 2. Call your Olympus customer service representative at 1-888-524-7266 option 1 to make arrangements to return your BF-Q180 bronchoscope and discuss replacement options. Olympus will issue a Return Material Authorization for you to return any BF-Q180 at no charge. 3. The timing to return your BF-Q180 and receive a replacement product from Olympus will depend on existing inventory. An Olympus representative will inform you of the approximate timing for your return/replacement. 4. In recognition of the need to continue to serve patients, the BF-Q180 may be continued to be used clinically until suitable alternatives are sufficiently available. Olympus recommends continued adherence to the Instructions for Use related to reprocessing. Customers should note that the BF-Q180 has been validated for High Level Disinfection (HLD), Ethylene Oxide Sterilization and Sterrad NX. Clinical support staff is available to answer questions related to all reprocessing procedures. 5. Olympus will reimburse you $10,000 for each returned BF-Q180. 6. Access the Olympus recall portal to indicate that you have received this notification. Go to https ://olympusamerica.com/recall. Enter the recall number Recall-0366 and provide your contact information as indicated in the portal. 7. If you may have further distributed the BF-Q180, please identify your customers, notify them at once of this product recall, and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this recall notification letter.
Quantity
5,813 devices distributed worldwide, including 2,648 devices distributed throughout U.S.A.
Distribution
Worldwide distribution - US Nationwide and International.
Lot, serial or model codes
All serial numbers
510(k) numbers
["K061313"]
PMA numbers
not on file
Product code
EOQ
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register