Skip to the content

FDA Medical Device Recalls (openFDA)

Canon Medical System, USA, INC.: Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Recall number
Z-3014-2026
Recalling firm
Canon Medical System, USA, INC.
Product
Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Status
Open, Classified
Initiated
2026-08-04
Posted
2026-08-25
Terminated
not on file
Reason
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Root cause
Process control
Action
On 08/04/2026, an "URGENT: MEDICAL DEVICE CORRECTION" letter was sent via email informing customers that, It has been found there is a possibility that the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose. Indications that the screws have loosened are abnormal noises and sensor errors from the ceiling mounted C-arm. Customers are instructed to: Should the suspended device make an abnormal noise, or a sensor error is displayed before the corrective action is implemented on your system, please stop using the device and call your local Canon Service Team. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. A representative of Canon Medical Systems USA will contact customers to schedule a time to check for the tightening and proper attachment of the fixing screws to correct this issue your system(s). For Questions - contact your service representative contact the InTouch team at intouchdispatch@us.medical.canon.
Quantity
10 systems
Distribution
U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
Lot, serial or model codes
Model Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: Pending SID Number: 30059180 Serial Number: STA2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30095017 Serial Number: SZA2432003 SID Number: 30097052 Serial Number: SZA2422002
510(k) numbers
["K182415","K203551","K210900"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register