Skip to the content

FDA Medical Device Recalls (openFDA)

Midmark Corporation: Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

Recall number
Z-3012-2026
Recalling firm
Midmark Corporation
Product
Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
Status
Open, Classified
Initiated
2026-07-20
Posted
2026-08-25
Terminated
not on file
Reason
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
Root cause
Software design
Action
n URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/17/26 was sent to customers. Safety Instructions Until the corrected IQecg Plugin software has been installed, for electrocardiogram tracings that reflect a fast heart rate (tachycardia), Midmark recommends performing a manual clinical assessment of the patient's heart rate prior to initiating therapeutic intervention. This bedside verification should be used to confirm the accuracy of the displayed ECG heart rate before making clinical management decisions based on the device's displayed value. If this issue is encountered, follow the temporary workaround instructions provided in Appendix A. Midmark will contact your facility with additional information regarding verification of your software installation and implementation of this Field Corrective Action. Product Correction Midmark has developed a corrected version of the IQecg Plugin software that addresses the issue described in this notification and is implementing a Field Corrective Action for affected software versions. No immediate action is required to contact Midmark at this time. Midmark will contact your facility to verify the IQecg Plugin software version(s) installed on your workstations and coordinate the appropriate correction process. Depending on your facility's configuration and support requirements, the correction may be completed using Midmark-provided installation instructions or with assistance coordinated by Midmark Technical Support or an authorized service provider. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email techsupport@midmark.com.
Quantity
3624 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot, serial or model codes
Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901
510(k) numbers
["K103640"]
PMA numbers
not on file
Product code
DPS
Firm FEI
1523530
Firm city
Versailles
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register