Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Axiom Artis U mobile x-ray system Product Usage: Mobile x-ray system

Recall number
Z-3009-2011
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Axiom Artis U mobile x-ray system Product Usage: Mobile x-ray system
Status
Terminated
Initiated
2011-06-29
Posted
2011-08-11
Terminated
2014-01-24
Reason
When the heat unit of the angiography system reaches or exceeds 65%, the cool down time display will appear with an acoustic signal. If users ignore the warning and continue the radiation, the tube may be autoprotected when it reaches its temperature limitation causing the X-Ray to lock until the tube cools down. Following a tube auto-protection, an error message "Door is Open" or "Err 421/240:
Root cause
Labeling False and Misleading
Action
Siemens sent a "Customer Safety Advisory Notice" letters to all affected users dated June 13, 2011. The letter identified the product, the problem, and the actions to be taken. Siemens included in the notice an addendum to the instruction for use to each system and request customers to take special attention. Customers were instructed to place the Safety Notice in the System's instruction for use. All appropriate personnel should be notified of the correction notice. If product was distributed further, customers were to forward the safety notice to the new owner and inform the firm of the new owner.
Quantity
18
Distribution
Nationwide Distribution (USA) including the states of: CA, CT, FL, KS, MN, MO, NJ, NY, NC, PA, and UT.
Lot, serial or model codes
Model number 10094012; serial numbers 20202, 20223, 20231, 20422, 20402, 20435, 21022, 22002, 21005, and 20432.   Model number 7008530; serial numbers 20146, 21112, 20003, 20115, 20135, 20006, 20113, and 20120.
510(k) numbers
["K040675"]
PMA numbers
not on file
Product code
IZL
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register