Teledyne Instruments - Analytical Instruments: Micro-Fuel Cell Oxygen Sensor, Class RI5 Sensor Part Number: C41131-R15 Teledyne Instruments - Analytical Instruments Disposable oxygen sensors with limited useful life. These sensors are accessories to other oxygen diagnostic devices, and are not in themselves suitable for therapeutic use.

Recall number
Z-3001-2011
Recalling firm
Teledyne Instruments - Analytical Instruments
Product
Micro-Fuel Cell Oxygen Sensor, Class RI5 Sensor Part Number: C41131-R15 Teledyne Instruments - Analytical Instruments Disposable oxygen sensors with limited useful life. These sensors are accessories to other oxygen diagnostic devices, and are not in themselves suitable for therapeutic use.
Status
Terminated
Initiated
2010-11-29
Posted
2011-08-11
Terminated
2011-08-22
Reason
The recall was initiated because during a test on a batch of sensors on the company shipping dock, Teledyne Instruments discovered electrolyte loss from the sealing cylinder on a few sensors. This test was done pursuant to a complaint from one customer of label discoloration caused because of the electrolyte loss. The risk of a leak form the oxygen sensor while in use is the disruption of therape
Root cause
Process design
Action
Teledyne Instruments sent emails and "Urgent--Device Field Corrections/Removal" letters in December 2010 to all affected customers. Th letter includes description of device and problem. Customers are advised to check them before use for any signs of electrolyte loss and if discoloration is noted to discontinue its use. and provides recommendation. Teledyne-desplaines to generate a product field corrections/removal log that lists names/addresses of the consignees, quantity of sensors shipped, sensor serial numbers and the date of shipment. For additional information or questions contact Teledyne Instruments at (626) 934-1504.
Quantity
2813
Distribution
Worldwide Distribution: Nationwide (US) distribution including the states of CA, TN and WI; and the countries of , Canada, Brazil, Chile, Saudi Arabia, and the U.K.
Lot, serial or model codes
Sensor Part Number: C41131-R15. Serial Numbers: 593875 to 593894 (20), 599156 to 599254(99), 611100 to 611599 (500), 617125 to 617674(550), 642023 to 642472 (450), 642474 to 642475 (2), 642485 to 643022 (538), 666158 to 666657 (500), 668499 to 668647 (149), and 668652 to 668656 (5).
510(k) numbers
["K001095"]
PMA numbers
not on file
Product code
not on file
Firm FEI
2026424
Firm city
City Of Industry
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register