Skip to the content

FDA Medical Device Recalls (openFDA)

ArthroCare Corporation: Smith&Nephew SmartStitch PerfectPasser Connector, REF # OM-8010, Rx Only, CE 0123, UDI: 00817470006971

Recall number
Z-3000-2020
Recalling firm
ArthroCare Corporation
Product
Smith&Nephew SmartStitch PerfectPasser Connector, REF # OM-8010, Rx Only, CE 0123, UDI: 00817470006971
Status
Terminated
Initiated
2019-02-21
Posted
not on file
Terminated
2022-08-17
Reason
The voluntary recall is being initiated due to a manufacturing error resulting in incorrect upper jaw dimensions, and the upper jaw could potentially disengage or detach during use.
Root cause
Device Design
Action
On February 21, 2019, the firm sent an "Urgent Medical Device Recall Notice" via Federal Express and E-Mail to all affected consignees. In addition to providing information on the recall product the firm asked consignees to take the following actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. 6. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Quantity
15,559 units
Distribution
US: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, SC, SG, TN, TX, UT, VA, WA, WI, WV OUS: Austria, Belgium, Canada, Switzerland, Germany, Denmark, France, England, Ireland, India, Netherlands, Norway, Poland, Puerto Rico, Russia, Singapore, Turkey, United Arab Emirates, South Africa)
Lot, serial or model codes
Batch Numbers: 1154857, 1179855, 1184077, 1194208, 2005153, 1154858, 1179856, 1184078, 1194763, 2005154, 1156429, 1179857, 1184825, 2001517, 2005251, 1165209, 1179858, 1184826, 2001518, 2005252, 1166194, 1179859, 1184827, 2001519, 2005494, 1168946, 1181657, 1184828, 2002828, 2006014, 1170928, 1181658, 1184829, 2002829, 2006037, 1170929, 1182577, 1184830, 2002830, 2006038, 1170930, 1182578, 1184831, 2002831, 2006039, 1170931, 1182579, 1185959, 2002832, 2006040, 1174530, 1182580, 1187253, 2004741, 1174531, 1182581, 1187254, 2004933, 1175357, 1182621, 1187255, 2004934, 1175401, 1182622, 1192014, 2005147, 1178577, 1184074, 1192015, 2005149, 1178578, 1184075, 1192016, 2005150, 1179854, 1184076, 1192017, 2005152
510(k) numbers
["K123268"]
PMA numbers
not on file
Product code
GAT
Firm FEI
3010813691
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register