FDA Medical Device Recalls (openFDA)
ArthroCare Corporation: Smith&Nephew SmartStitch PerfectPasser Connector, REF # OM-8010, Rx Only, CE 0123, UDI: 00817470006971
- Recall number
- Z-3000-2020
- Recalling firm
- ArthroCare Corporation
- Product
- Smith&Nephew SmartStitch PerfectPasser Connector, REF # OM-8010, Rx Only, CE 0123, UDI: 00817470006971
- Status
- Terminated
- Initiated
- 2019-02-21
- Posted
- not on file
- Terminated
- 2022-08-17
- Reason
- The voluntary recall is being initiated due to a manufacturing error resulting in incorrect upper jaw dimensions, and the upper jaw could potentially disengage or detach during use.
- Root cause
- Device Design
- Action
- On February 21, 2019, the firm sent an "Urgent Medical Device Recall Notice" via Federal Express and E-Mail to all affected consignees. In addition to providing information on the recall product the firm asked consignees to take the following actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. 6. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
- Quantity
- 15,559 units
- Distribution
- US: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, SC, SG, TN, TX, UT, VA, WA, WI, WV OUS: Austria, Belgium, Canada, Switzerland, Germany, Denmark, France, England, Ireland, India, Netherlands, Norway, Poland, Puerto Rico, Russia, Singapore, Turkey, United Arab Emirates, South Africa)
- Lot, serial or model codes
- Batch Numbers: 1154857, 1179855, 1184077, 1194208, 2005153, 1154858, 1179856, 1184078, 1194763, 2005154, 1156429, 1179857, 1184825, 2001517, 2005251, 1165209, 1179858, 1184826, 2001518, 2005252, 1166194, 1179859, 1184827, 2001519, 2005494, 1168946, 1181657, 1184828, 2002828, 2006014, 1170928, 1181658, 1184829, 2002829, 2006037, 1170929, 1182577, 1184830, 2002830, 2006038, 1170930, 1182578, 1184831, 2002831, 2006039, 1170931, 1182579, 1185959, 2002832, 2006040, 1174530, 1182580, 1187253, 2004741, 1174531, 1182581, 1187254, 2004933, 1175357, 1182621, 1187255, 2004934, 1175401, 1182622, 1192014, 2005147, 1178577, 1184074, 1192015, 2005149, 1178578, 1184075, 1192016, 2005150, 1179854, 1184076, 1192017, 2005152
- 510(k) numbers
- ["K123268"]
- PMA numbers
- not on file
- Product code
- GAT
- Firm FEI
- 3010813691
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28