Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline medical convenience kits, labeled as: 1) NEURO CDS, Medline Kit SKU CDS985246J; 2) NEURO PACK, Medline Kit SKU DYNJ32770V; 3) NEURO BUNDLE-LF, Medline Kit SKU DYNJ40095D; 4) NEURO PACK-LF, Medline Kit SKU DYNJ42844D; 5) NEURO PACK, Medline Kit SKU DYNJ45121I; 6) NEURO LAMINECTOMY PACK, Medline Kit SKU DYNJ58012A; 7) NEURO PACK SJ, Medline Kit SKU DYNJ63477A; 8) BASIC NEURO PACK, Medline Kit SKU DYNJ83071A.

Recall number
Z-2996-2026
Recalling firm
Medline Industries, LP
Product
Medline medical convenience kits, labeled as: 1) NEURO CDS, Medline Kit SKU CDS985246J; 2) NEURO PACK, Medline Kit SKU DYNJ32770V; 3) NEURO BUNDLE-LF, Medline Kit SKU DYNJ40095D; 4) NEURO PACK-LF, Medline Kit SKU DYNJ42844D; 5) NEURO PACK, Medline Kit SKU DYNJ45121I; 6) NEURO LAMINECTOMY PACK, Medline Kit SKU DYNJ58012A; 7) NEURO PACK SJ, Medline Kit SKU DYNJ63477A; 8) BASIC NEURO PACK, Medline Kit SKU DYNJ83071A.
Status
Open, Classified
Initiated
2026-07-08
Posted
2026-08-21
Terminated
not on file
Reason
Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges.
Root cause
Nonconforming Material/Component
Action
Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 7/8/2026 via USPS and email. The notice explained the issue, potential adverse health consequences, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-112-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
3468 kits
Distribution
US Nationwide distribution in the states of AZ, CA, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NY, OH, TX, WI, WY.
Lot, serial or model codes
Medline Kit SKU CDS985246J, UDI/DI 10193489232394 (each) 40193489232395 (case), Lot Number 22HMB295; Medline Kit SKU CDS985246J, UDI/DI 10193489232394 (each) 40193489232395 (case), Lot Number 22IMB860; Medline Kit SKU DYNJ32770V, UDI/DI 10195327090111 (each) 40195327090112 (case), Lot Number 22FMB806; Medline Kit SKU DYNJ40095D, UDI/DI 10195327281250 (each) 40195327281251 (case), Lot Number 23GMI744; Medline Kit SKU DYNJ42844D, UDI/DI 10193489880564 (each) 40193489880565 (case), Lot Number 25EBV362; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25CMH980; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25FMB096; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25GMA066; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25GMH940; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25BMI081; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25IMC735; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24DME519; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24EMG256; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24GMJ491; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24IMA243; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24JMA162; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24JMD784; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24KMI470; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24ABN077; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24ABN475; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24DBA116; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23AMH395; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23EMA407; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23HBH164; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23HBO192; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23HBV863; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23IBF802; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23IBS028; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23JBV613; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22BME230; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22CMA406; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22EMB248; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22EMI134; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22HME564; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22IMA455; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22JMI068; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22LMB601; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 24BMI624; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 24EME751; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23AMJ022; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23CMB364; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23EMD743; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23FMH669; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23JME474; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 22DMH314; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 22LME727; Medline Kit SKU DYNJ63477A, UDI/DI 10195327123451 (each) 40195327123452 (case), Lot Number 22EMB958; Medline Kit SKU DYNJ63477A, UDI/DI 10195327123451 (each) 40195327123452 (case), Lot Number 22EMC777; Medline Kit SKU DYNJ83071A, UDI/DI 10195327452490 (each) 40195327452491 (case), Lot Number 23HMF326.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OJG
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register