Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: NM 830 Model # H3910AC

Recall number
Z-2995-2020
Recalling firm
GE Healthcare, LLC
Product
NM 830 Model # H3910AC
Status
Terminated
Initiated
2020-08-14
Posted
not on file
Terminated
2021-05-06
Reason
Rotor bearing screws were found loose on detector.
Root cause
Use error
Action
GE Healthcare issued an Urgent Medical device correction notification on August 14, 2020. It states the following: 1. GE Healthcare will inspect and, if required, correct all affected products at no cost to you. 2. A GE Healthcare representative will contact you to arrange for the correction 3. Please complete the acknowledgement form and return it to GE Healthcare promptly upon receipt and no later than 30 days from receipt to nm.fmi40890.responses@ge.com If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
25 units
Distribution
USA : AR CT FL GA IL KS KY LA MA MD MI MN MO NE NJ NV NY OH OK RI SC TN TX VA VT WA WI OUS: ARGENTINA AUSTRIA BELGIUM BRAZIL Canada CHINA FRANCE GERMANY GREECE INDIA ISRAEL ITALY JAPAN Korea (Republic of) NETHERLANDS NORWAY PANAMA POLAND PORTUGAL Russia SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED KINGDOM VIETNAM
Lot, serial or model codes
Serial # 830X60001 System ID: 479338N830 Serial # 830X60004 System ID: 901227NM1 Serial # 830X60018 System ID: 772468NM830 Serial # 830X60012 System ID: 813315D830 Serial # 830X60011 System ID: 225761NM1 Serial # 830X60014 System ID: 225761NM Serial # 830X60016 System ID: 225761NM Serial # 830X60019 System ID: 573458NM Serial # 830X60023 System ID: 718238D830 Serial # 830X60006 System ID: 716862NM830 Serial # 830X60015 System ID: 585396NM Serial # 830X60003 System ID: 585786NM830 Serial # 830X60025 System ID: 405717NM830 Serial # 830X60021 System ID: 615316NM830 Serial # 830X60017 System ID: 956323NM830 Serial # 830X60002 System ID: 100039NU05 Serial # 830X60022 System ID: 705743DNM830 Serial # 830X60024 System ID: 416898NM830 Serial # 830X60013 System ID: M4040557 Serial # 830X60010 System ID: NS0102 Serial # 830X60020 System ID: NS0101 Serial # 830X60009 System ID: 0850260168 Serial # 830X60005 System ID: 0850260165 Serial # 830X60008 System ID: 0850260166 Serial # 830X60007 System ID: 752030NU01
510(k) numbers
["K111445"]
PMA numbers
not on file
Product code
KPS
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register